SYRINGE 5ML HEPARIN 10 UNIT
Received Apr 29, 2025 · Event occurred Apr 19, 2025
Report 1911916-2025-00315 · MDR key 21926439
Device
Generic name
Heparin, Vascular Access Flush
Manufacturer
Becton DickinsonCatalog number
306414
Lot number
434180N
Product problems
- Infusion or Flow Problem
- Infusion or Flow Problem
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
(B)(4) FOLLOW UP: IT WAS REPORTED THE SYRINGE HAD AIR IN IT AND LACKED HEPARIN FLUID. OUR QUALITY TEAM HAS COMPLETED A DEVICE HISTORY RECORD REVIEW FOR MATERIAL NUMBER 306414 AND LOT NUMBER 434180N. THE REVIEW DID NOT IDENTIFY ANY ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THE DEFECT, AND ALL QUALITY TESTS WERE WITHIN SPECIFICATION. DUE TO THE UNAVAILABILITY OF A SAMPLE FOR RETURN, A COMPREHENSIVE SAMPLE INVESTIGATION COULD NOT BE CONDUCTED. SHOULD YOU ENCOUNTER ANY FURTHER ISSUES WITH OUR PRODUCT, WE WOULD APPRECIATE THE OPPORTUNITY TO PERFORM A DETAILED ANALYSIS.
Additional Manufacturer Narrative
INITIAL MDR SUBMISSION. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED.
Description of Event or Problem
MATERIAL #306414 BATCH # 434180N IT WAS REPORTED BY THE CUSTOMER THAT THE SYRINGE LACKED HEPARIN/FLUID. VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. EMAIL(S) ATTACHED. SHARING TWO ADDITIONAL EVENTS OF AIR IN SYRINGE AND LACK OF HEPARIN/FLUID. THESE WERE NOT SAVED FOR RETURN. ASKING FOR ADDITIONAL INVESTIGATION INTO THE MANUFACTURING DEFECT. THE CLOSURE LETTERS I RECEIVED FOR THE PRIOR EVENTS STATE NO ISSUES WITH THE PRODUCTION PROCESS BUT THIS IS AN INCREASING SAFETY TREND FOR THIS PRODUCT. EVENT DATE: (B)(6) 2025. REF (B)(4). LOT 434180N. EXP 2026-06-30. EVENT DATE: (B)(6) 2025. REF (B)(4). LOT 43410N. EXP 2026-06-30.