ALINITY I CA 19-9XR REAGENT KIT
Received Apr 28, 2025 · Event occurred Apr 8, 2025
Report 3002809144-2025-00146 · MDR key 21923511
Device
Generic name
System, Test, Carbohydrate Ag (ca19-9), For Monit And Mgmt Of Pancreatic Cancer
Manufacturer
Abbott GmbhCatalog number
08P32-30
Lot number
68329FP00
Product problems
- High Test Results
- High Test Results
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 08P32-20 THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 8P32-21 / 31, WITH 510K/PMA/BLA NUMBER K052000. AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.
Description of Event or Problem
THE CUSTOMER OBSERVED FALSELY ELEVATED ALINITY I CA 19-9XR AND PROVIDED THE FOLLOWING DATA: EARLY MARCH: AUTOMATIC DILUTION WITH ALINITY WAS 2,200U/ML. END OF MARCH: AUTOMATIC DILUTION WITH ALINITY: AROUND 1,500U/ML, GOOD REPRODUCIBILITY. OUTSOURCED: FUJIREBIO LUMIPULSE PLATFORM GENERATED A RESULT OF 15U/ML (NEGATIVE). THE CUSTOMER REPORTED LOTS 70427FP00 AND 68329FP00 WERE IMPACTED, HOWEVER DID NOT PROVIDE WHICH RESULTS CAME FROM WHICH LOT. ADDITIONAL LABORATORY AND PATIENT DATA: MEDICAL HISTORY/DIAGNOSIS: UNKNOWN. DUPAN-2: NEGATIVE. CEA: NEGATIVE. NORMAL DIAGNOSTIC IMAGING. THE CUSTOMER WAS NOT ABLE TO PROVIDE ANY FURTHER PATIENT MEDICAL INFORMATION. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.
Description of Event or Problem
THE CUSTOMER OBSERVED FALSELY ELEVATED ALINITY I CA 19-9XR AND PROVIDED THE FOLLOWING DATA: EARLY MARCH: AUTOMATIC DILUTION WITH ALINITY WAS 2,200U/ML END OF MARCH: AUTOMATIC DILUTION WITH ALINITY: AROUND 1,500U/ML, GOOD REPRODUCIBILITY. OUTSOURCED: FUJIREBIO LUMIPULSE PLATFORM GENERATED A RESULT OF 15U/ML (NEGATIVE) THE CUSTOMER REPORTED LOTS 70427FP00 AND 68329FP00 WERE IMPACTED, HOWEVER DID NOT PROVIDE WHICH RESULTS CAME FROM WHICH LOT. ADDITIONAL LABORATORY AND PATIENT DATA: MEDICAL HISTORY/DIAGNOSIS: UNKNOWN. DUPAN-2: NEGATIVE. CEA: NEGATIVE. NORMAL DIAGNOSTIC IMAGING. THE CUSTOMER WAS NOT ABLE TO PROVIDE ANY FURTHER PATIENT MEDICAL INFORMATION. THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.
Additional Manufacturer Narrative
THE COMPLAINT INVESTIGATION INCLUDED A REVIEW OF DATA AND INFORMATION PROVIDED BY THE CUSTOMER, SEARCH FOR SIMILAR COMPLAINTS, TRENDING REVIEW, LABELING REVIEW, DEVICE HISTORY RECORD REVIEW, AND FIELD DATA REVIEW. RETURN TESTING WAS NOT COMPLETED AS RETURNS WERE NOT AVAILABLE. DATA AND INFORMATION PROVIDED BY THE CUSTOMER WERE REVIEWED AND SUPPORT THE COMPLAINT ISSUE WITHOUT INDICATION FOR ANY ADDITIONAL ISSUE. DEVICE HISTORY REVIEW FOR THE COMPLAINT LOT DID NOT IDENTIFY ANY ISSUES. A REVIEW OF TRACKING AND TRENDING DID NOT IDENTIFY AN INCREASE IN COMPLAINTS FOR THE ALINITY I CA 19-9XR AND DID NOT IDENTIFY AN INCREASE IN COMPLAINT ACTIVITY FOR THE COMPLAINT LOT. THE OVERALL PERFORMANCE OF THE ALINITY I CA 19-9XR REVIEWED USING FIELD DATA FROM CUSTOMERS. THE REVIEW OF FIELD DATA DETERMINED THE MEDIAN PATIENT RESULTS FOR LOT 68329FP00WAS WITHIN THE ESTABLISHED LIMITS. LABELING WAS REVIEWED AND WAS FOUND TO ADDRESS THE CUSTOMER REPORTED EVENT. BASED ON THE AVAILABLE INFORMATION, NO SYSTEMIC ISSUE OR DEFICIENCY WAS IDENTIFIED FOR THE ALINITY I CA 19-9XR REAGENT LOT 68329FP00.