6257 XPEDITION POWERED CHAIR
Received Apr 28, 2025 · Event occurred Jan 1, 2025
Report 0001831750-2025-00575 · MDR key 21919978
Device
Generic name
Transport, Patient, Powered
Manufacturer
Stryker Medical-kalamazooCatalog number
625700000000
Product problems
- Difficult or Delayed Positioning
- Difficult or Delayed Positioning
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. 2 DEVICES ARE PENDING EVALUATION. THERE WAS NO REMEDIAL ACTION TAKEN. THIS DEVICE IS NOT LABELED FOR SINGLE USE.
Description of Event or Problem
THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED DIFFICULT TO MANEUVER UNIT WHICH DOES NOT RESULT IN A TIP. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Description of Event or Problem
THIS REPORT NOW SUMMARIZES 1 MALFUNCTION EVENTS, INSTEAD OF 2.
Additional Manufacturer Narrative
1 DEVICE THAT WAS PENDING EVALUATION WAS EVALUATED IN THE FIELD AND THE ISSUE WAS CONFIRMED. THE DEVICE WAS REPAIRED ON SITE AND RETURNED TO SERVICE. 1 DEVICE THAT WAS PENDING WAS EVALUATED IN THE FIELD BUT THE ISSUE WAS NOT CONFIRMED; NO DEFECT OR MALFUNCTION WAS FOUND. SECTION H CODES HAVE BEEN UPDATED.