inforMED
MalfunctionJDQ

APPLICATION INSTRUMENT FOR SNAL ZIPFIX

Received Apr 28, 2025

Report 8030965-2025-04141 · MDR key 21919740

Device

Generic name

Fixation Cerclage

Manufacturer

Synthes Gmbh

Catalog number

03.501.080

Lot number

L974545

Product problems

  • Break
  • Break

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. H11 ADDITIONAL NARRATIVE: ADDED: D9 CORRECTED: G1. H3, H6: THE CUSTOMER REPORTED THE APPLICATION INSTRUMENT FOR SNAL ZIPFIX CUTTING LEVER BROKE OFF THE APPLICATOR GUN. THE REPAIR TECHNICIAN REPORTED BROKEN HANDLE . THE ITEM IS NOT REPAIRABLE PER THE INSPECTION SHEET. THE CAUSE OF THE ISSUE IS END OF LIFE. THE ITEM WILL BE FORWARDED TO CUSTOMER QUALITY. THE EVALUATION WAS CONFIRMED. DEVICE HISTORY REVIEW PART# 03.501.080 LOT # L974545 MANUFACTURING SITE: WERK HAGENDORF RELEASE TO WAREHOUSE DATE: 30 AUGUST 2018. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED ARTICLE LOT AND NO NON-CONFORMANCES WERE IDENTIFIED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.

Description of Event or Problem

IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE APPLICATION INSTRUMENT FOR SNAL ZIPFIX CUTTING LEVER BROKE OFF THE APPLICATOR GUN. IT IS UNKNOWN WHEN THE ISSUE WAS OBSERVED.

Additional Manufacturer Narrative

PRODUCT COMPLAINT # (B)(4). DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.