Received Apr 25, 2025 · Event occurred Apr 1, 2025

Report 9616086-2025-00043 · MDR key 21915074

Device

Generic name

Legend Xt 2ch Stim Pkg

Manufacturer

Djo Llc

Catalog number

2763

Lot number

T2687

Product problems

  • Adverse Event Without Identified Device or Use Problem
  • Adverse Event Without Identified Device or Use Problem

Patient

NA · Male

  • Burn(s)
  • Burn(s)

Narrative

Additional Manufacturer Narrative

IT WAS REPORTED THAT PATIENT SUSTAINED 2ND DEGREE BURNS UNDER 3 ELECTRODES DURING TREATMENT THAT REQUIRED MEDICAL TREATMENT AT URGENT CARE THE DEVICE HAS BEEN RETURNED AND EVALUATED. THE DEVICE PASSED ALL FUNCTIONAL TESTING AND IS IN GOOD SHAPE. LEAD WIRES WERE TESTED AND SHOW GOOD CONTINUITY. THE ROOT CAUSE NOT PROPERLY PREPARING SKIN OR USING OFF BRAND ACCESSORIES COULD CAUSE BURNS UNDER STIM THERAPY.

Description of Event or Problem

IT WAS REPORTED THAT PATIENT SUSTAINED 2ND DEGREE BURNS UNDER 3 ELECTRODES DURING TREATMENT THAT REQUIRED MEDICAL TREATMENT AT URGENT CARE.