COUPLER AC FFL 19
Received Apr 24, 2025 · Event occurred Jan 1, 2025
Report 1221934-2025-01481 · MDR key 21902042
Device
Generic name
Accessories, Photographic, For Endoscope (exclude Light Sources)
Manufacturer
Medos International Sã rlCatalog number
242437
Product problems
- Detachment of Device or Device Component
- Material Deformation
- Detachment of Device or Device Component
- Material Deformation
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THIS REPORT IS BEING SUBMITTED IN PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY MITEK OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, MITEK, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. H11 ADDITIONAL NARRATIVE: H4: UNKNOWN. AS OF THIS DATE, THE DEVICE HAS NOT BEEN RETURNED FOR EVALUATION; THEREFORE, THE REPORTED CONDITION CANNOT BE CONFIRMED AND/OR DUPLICATED.
Description of Event or Problem
IT WAS REPORTED THAT THE COUPLER AC FFL 19 DEVICE SCREW ON THE QUICK RELEASE CLAMP PART OF THE COUPLER HAS COME OFF. IT WAS NOT REPORTED IF THE DEVICE WAS USED IN SURGERY. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN A SURGICAL PROCEDURE, OR IF A SPARE DEVICE WAS AVAILABLE FOR USE. THE EXACT DATE OF THIS EVENT WAS UNKNOWN BUT WAS NOTED TO HAVE OCCURRED IN 2025. IT WAS NOT REPORTED IF THERE WERE ANY INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Additional Manufacturer Narrative
THIS REPORT IS BEING SUBMITTED IN PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY MITEK OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, MITEK, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. H11 ADDITIONAL NARRATIVE: INVESTIGATION SUMMARY: THE COMPLAINT DEVICE WAS RECEIVED AT THE SERVICE CENTER AND EVALUATED. DURING THE SERVICE EVALUATION THE FOLLOWING DEFECTS WERE IDENTIFIED: VISUAL: MISSING COMPONENTS. IT WAS CONFIRMED THAT THE SCREW OF THE QUICK RELEASE CLAMP IS DETACHED. PER SERVICE REPORTS, THIS COMPLAINT WAS CONFIRMED. BASED ON SERVICE MANUAL OPERATIONAL AND DIAGNOSTIC, THE DEVICE WAS FOUND TO BE NON-REPAIRABLE, AND IT WAS RETURNED UNREPAIRED TO THE CUSTOMER. BASED ON THE INVESTIGATION FINDINGS, THE POTENTIAL ROOT CAUSE IS TRACED TO FAULTY PARTS. IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED AND THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE REPORTED COMPLAINT CONDITION. AS PART OF J&J MEDTECH ORTHOPAEDICS QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES.