UNICEL® DXC 600I SYNCHRON® ACCESS® CLINICAL SYSTEM
Received Aug 3, 2011 · Event occurred Jul 13, 2011
Report 2050012-2011-03822 · MDR key 2190005
Device
Generic name
Analyzer, Chemistry
Manufacturer
Beckman Coulter, Inc.Model number
NACatalog number
A27318
Lot number
NA
Product problems
- Improper or Incorrect Procedure or Method
- High Test Results
Patient
Not reported
- No Code Available
Narrative
Additional Manufacturer Narrative
USER ERROR CONTRIBUTED OR CAUSED THIS EVENT. NO SERVICE WAS DISPATCHED TO THE SITE FOR THIS EVENT. MDRS ASSOCIATED WITH THIS EVENT: 2050012-2011-03822 AND 2050012-2011-03823.
Description of Event or Problem
THE CUSTOMER REPORTED THAT ERRONEOUSLY HIGH TOTAL BILIRUBIN (TBIL) RESULTS WERE GENERATED FROM A UNICEL DXC 600I SYNCHRON ACCESS CLINICAL SYSTEM FOR ONE PATIENT'S SAMPLES OVER TWO DAYS. THIS REPORT IS ONE OF TWO AND REPRESENTS THE ERRONEOUSLY HIGH TOTAL BILIRUBIN (TBIL) RESULT GENERATED FROM A UNICEL DXC 600I SYNCHRON ACCESS CLINICAL SYSTEM ON (B)(6) 2011 FOR ONE PATIENT SAMPLE. THE INITIAL TBIL RESULT WAS A SUPPRESSED VALUE WITH AN "OUT OF INSTRUMENT RANGE HIGH" INSTRUMENT FLAG. UPON REPEAT TESTING OF A DILUTION OF THE SAMPLE, THE LABORATORY TECHNICIAN INCORRECTLY CALCULATED THE END RESULT. THE TECHNICIAN DID NOT FOLLOW THE CALCULATION INSTRUCTIONS PROVIDED IN A BECKMAN COULTER INC. TECHNICAL APPLICATIONS LETTER DATED (B)(6) 2007. DUE TO THIS ERROR, THE REPORTED RESULT WAS TEN TIMES THE CORRECT RESULT. THE PATIENT WAS ADMITTED TO THE HOSPITAL. IT IS UNKNOWN AS TO WHETHER HOSPITAL ADMISSION WAS BASED UPON THIS INCORRECT RESULT, HOWEVER FOR THE PURPOSES OF THIS REPORT IT IS ASSUMED THAT HOSPITAL ADMISSION WAS LINKED TO THE ERRONEOUS TBIL RESULT. A PHYSICIAN QUESTIONED THE HIGH RESULT AND THE SAMPLE WAS RETESTED ON A DILUTED SAMPLE UTILIZING THE CORRECT CALCULATION TO GENERATE THE FINAL RESULT. THIS RESULT WAS LOWER, MORE BELIEVABLE, AND REGARDED AS VALID. SPECIFIC PATIENT INFORMATION AND SAMPLE COLLECTION/HANDLING INFORMATION WAS NOT SUPPLIED BY THE CUSTOMER. INSTRUMENT TBIL QUALITY CONTROLS RESULTS DURING THE TIMEFRAME OF THE EVENT WERE FOUND TO BE WITHIN SPECIFICATION.