OXY2MASK O2/ETCO2 ADULT
Received Apr 22, 2025 · Event occurred Apr 4, 2025
Report 21881996 · MDR key 21881996
Device
Generic name
Analyzer, Gas, Carbon-dioxide, Gaseous-phase
Manufacturer
Southmedic, IncModel number
OM-2125-14SLMCatalog number
OM-2125-14SLM
Lot number
W1005506
Product problems
- Incorrect Measurement
- Incorrect, Inadequate or Imprecise Result or Readings
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
PER PROTOCOL, WE ARE SUPPOSED TO MONITOR ETCO2 [END-TIDAL CARBON DIOXIDE] ON ALL OF OUR PATIENTS WITH OSA [OBSTRUCTIVE SLEEP APNEA] WHO RECEIVE MODERATE SEDATION. WE CURRENTLY USE THE O2/ETCO2 ADULT OXYMASK TO BOTH PROVIDE OXYGEN AND MEASURE ETCO2. REFERENCE NUMBER OM-2125-14SLM. MULTIPLE NURSES HAVE HAD DIFFICULTY MEASURING ACCURATE CO2 [CARBON DIOXIDE] LEVELS ON OUR SEDATED PATIENTS, WHICH WE BELIEVE IS DUE TO THE EXCESSIVE LENGTH OF THE CO2 TUBING. WE ARE ALL GETTING READINGS BETWEEN 7-11 MMHG, WHEN THEY SHOULD BE 35-45 MMHG.