MalfunctionNYU
PALINDROME
Received Apr 22, 2025 · Event occurred Apr 7, 2025
Report 21881353 · MDR key 21881353
Device
Generic name
Catheter, Hemodialysis, Implanted, Coated
Manufacturer
Mozarc Medical Us LlcModel number
8888123405PCatalog number
8888123405P
Lot number
2412000239
Product problems
- Failure to Advance
- Defective Device
Patient
58 YR · Male
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
WHILE PERFORMING A HD [HEMODIALYSIS] CATHETER EXCHANGE IT WAS NOTICED THAT THE STIFFENERS INSIDE THE KIT DID NOT HAVE HOLES THAT ALLOW FOR A WIRE TO PASS THROUGH AND ADVANCE THE CATHETER INTO POSITION. ADDITIONAL SUPPLIES WERE NEEDED SO PROCEDURE WAS COMPLETED WITHOUT ISSUE. SUPPLY WAS A 23CM PALINDROME HEPARIN COATED HEMODIALYSIS CATHETER - REF#: 8888123405P - LOT#: 2412000239.