AFINION 2
Received Apr 22, 2025 · Event occurred Apr 4, 2025
Report 9613069-2025-00003 · MDR key 21881228
Device
Generic name
Analyzer
Manufacturer
Abbott Diagnostics Technologies AsModel number
1116770Catalog number
1116770
Lot number
10218683
Product problems
- Smoking
- Smoking
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
ON (B)(6) 2025 AN END-USER IN SOUTH KOREA PHONED TECHNICAL SERVICES (TS) TO REPORT THAT SMOKE AND A BURNING SMELL WAS COMING FROM AN AFINION 2 INSTRUMENT (SN NOT PROVIDED). THE END-USER TURNED OFF THE INSTRUMENT FOR THE SAFETY ISSUE AND WERE ADVISED BY TS NOT TO SWITCH ON AGAIN. TS CONFIRMED WITH THE END USER THAT THERE WAS NO INJURIES OR HARM. NO PATIENT INVOLVED AND NO IMPACT TO PATIENT. ON (B)(6) 2025 THE ABBOTT REPRESENTATIVE CALLED TS AND PROVIDED THE SERIAL NUMBER OF THE AFINION 2 INSTRUMENT (SN: (B)(6)).
Additional Manufacturer Narrative
THE EVENT IS FOR SIMILAR PRODUCT SOLD IN USA, THE UDI PROVIDED IN SECTION D4 IS THE ACTUAL PRODUCT AND NOT USA VARIANT OF DEVICE. ON (B)(6) 2025. THE ABBOTT REPRESENTATIVE ALSO CONFIRMED THAT THE FOLLOWING FOLLOW-UP ACTIONS HAD BEEN CARRIED OUT BY THE DISTRIBUTOR: - DISTRIBUTOR HAD RETRIEVED THE INSTRUMENT FROM THE END-USER AND TESTED IT FOR OVER TWO HOURS BUT DID NOT ENCOUNTER ANY ISSUES. - PROVIDED A REPLACEMENT INSTRUMENT AS THE END-USER REPORTED SMOKE AND A BURNING SMELL COMING ON (B)(6) 2025, TS INITIATED A RETURN REQUEST, FOR THE AFINION 2 TO BE RETURNED TO THE LEGAL MANUFACTURER, ABBOTT DIAGNOSTICS TECHNOLOGIES AS.
Description of Event or Problem
ON (B)(6) 2025 AN END-USER IN SOUTH KOREA PHONED TECHNICAL SERVICES (TS) TO REPORT THAT SMOKE AND A BURNING SMELL WAS COMING FROM AN AFINION 2 INSTRUMENT (SN NOT PROVIDED). THE END-USER TURNED OFF THE INSTRUMENT FOR THE SAFETY ISSUE AND WERE ADVISED BY TS NOT TO SWITCH ON AGAIN. TS CONFIRMED WITH THE END USER THAT THERE WAS NO INJURIES OR HARM. NO PATIENT INVOLVED AND NO IMPACT TO PATIENT. ON 10 APRIL 2025 THE ABBOTT REPRESENTATIVE CALLED TS AND PROVIDED THE SERIAL NUMBER OF THE AFINION 2 INSTRUMENT (SN: (B)(6)).
Additional Manufacturer Narrative
THE EVENT IS FOR SIMILAR PRODUCT SOLD IN USA, THE UDI PROVIDED IN SECTION D4 IS THE ACTUAL PRODUCT AND NOT USA VARIANT OF DEVICE. ABBOTT REPRESENTATIVE CONFIRMED THAT THE FOLLOWING FOLLOW-UP ACTIONS HAVE BEEN CARRIED OUT BY THE DISTRIBUTOR: DISTRIBUTOR RETRIEVED THE INSTRUMENT FROM THE END-USER AND TESTED IT FOR OVER TWO HOURS BUT DID NOT ENCOUNTER ANY ISSUES. HOWEVER, SINCE THE END-USER REPORTED SMOKE AND A BURNING SMELL COMING FROM THE INSTRUMENT, A REPLACEMENT WAS PROVIDED AS A PRECAUTION. THIS ACTION WAS ALSO COMMUNICATED TO THE SUB-DEALER. THE USER MANUAL EXPLAINS TO PLACE YOUR AFINION 2 ANALYZER ON A DRY, CLEAN, STABLE AND HORIZONTAL SURFACE. MAKE SURE THAT THE ANALYZER IS LOCATED WITH SUFFICIENT SURROUNDING AIRSPACE, AT LEAST 10 CM ON EACH SIDE. PLACEMENT OF AFINION 2 ANALYZER SHOULD ALLOW EASY DISCONNECTION FROM THE WALL OUTLET AT ANY TIME. ACCLIMATE THE ANALYZER TO AMBIENT OPERATING TEMPERATURE (15-32°C, 59-89°F) BEFORE USE. THE ANALYZER MIGHT BE IMPAIRED BY: ¿ CONDENSING HUMIDITY AND WATER. ¿ HEAT AND LARGE TEMPERATURE VARIATIONS. ¿ DIRECT SUNLIGHT. ¿ VIBRATIONS (E.G. FROM CENTRIFUGES AND DISHWASHERS). ¿ ELECTROMAGNETIC RADIATION. ¿ MOVEMENT OF THE ANALYZER DURING PROCESSING OF A TEST CARTRIDGE. THE COMPLAINT DETAILS WERE REVIEWED ALONG WITH OUR INTERNAL RECORDS. THE INTERNAL INVESTIGATION CONCLUDED THAT THE SMOKE AND BURNING ODOR REPORTED IN CONNECTION WITH THE INSTRUMENT WERE ATTRIBUTABLE TO A DEFORMED LID MOTOR BRACKET. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE REPORTED IN THE COMPLAINT. CUSTOMER COMPLAINTS WILL CONTINUE TO BE MONITORED TO ENSURE CONTROL LIMITS ARE NOT EXCEEDED AND ENSURE PRODUCT'S PERFORMANCE MEETS CUSTOMERS' EXPECTATIONS. THANK YOU FOR TAKING THE TIME TO BRING THIS EVENT TO OUR ATTENTION. IF WE CAN PROVIDE ANY FURTHER ASSISTANCE, PLEASE DO NOT HESITATE TO CONTACT US.
Remedial action
- Replace