EARLYVUE VS30 VITAL SIGNS MONITOR
Received Apr 22, 2025 · Event occurred Mar 27, 2025
Report 9610816-2025-000332 · MDR key 21880902
Device
Generic name
Earlyvue Vs30 Vital Signs Monitor
Manufacturer
Philips Medizin Systeme Böblingen GmbhModel number
863359Catalog number
863359
Product problems
- High Readings
- High Readings
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
PHILIPS RECEIVED A COMPLAINT ON THE EARLYVUE VS30 VITAL SIGNS MONITOR INDICATING THE SYSTEM HAS HIGH SATURATION READINGS. IT IS UNKNOWN IF THE DEVICE WAS IN USE, MONITORING A PATIENT AT THE TIME OF THE EVENT. NO ADVERSE EVENT INVOLVING A PATIENT OR USER WAS REPORTED.
Additional Manufacturer Narrative
PHILIPS IS IN THE PROCESS OF OBTAINING ADDITIONAL INFORMATION CONCERNING THIS EVENT AND THE COMPLAINT IS STILL UNDER INVESTIGATION. A FINAL REPORT WILL BE SUBMITTED ONCE THE INVESTIGATION IS COMPLETE.
Additional Manufacturer Narrative
DIAGNOSTIC/FUNCTIONAL TESTING WAS PERFORMED AT THE PHILIPS AUTHORIZED REPAIR FACILITY. RESULTS OF FUNCTIONAL TESTING INDICATE THAT THE DEVICE WORKED AS DESIGNED. THE DEVICE'S OXYGEN SATURATION WAS CHECKED AND WAS WITHIN THE ESTABLISHED VALUES.