NRFIT EXTENSION SET W/ FILTER
Received Apr 17, 2025 · Event occurred Jan 11, 2025
Report 3012307300-2025-04514 · MDR key 21860844
Device
Generic name
Administrations Sets With Neuraxial Connectors
Manufacturer
Smiths Healthcare Manufacturing S.a. De C.v.Catalog number
21-7608-24
Lot number
4405868
Product problems
- Fluid/Blood Leak
- Fluid/Blood Leak
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THIS FILE WAS ORIGINALLY INCORRECTLY FILED UNDER REGISTRATION NUMBER MRN 9617604-2025-00118. THE DATE OF THAT SUBMISSION WAS 2/10/2025. DEVICE EVALUATION: ONE DEVICE WAS RECEIVED FOR DEVICE ANALYSIS. FUNCTIONAL TESTING AND VISUAL INSPECTION WAS PERFORMED. VISUAL INSPECTION FOUND NO DAMAGE OR ANOMALIES. FUNCTIONAL TESTING COULD CONFIRM THE REPORTED CUSTOMER COMPLAINT. A LEAK WAS OBSERVED FROM THE FILTER OUTLET TUBE. THE ROOT CAUSE IS DUE TO INCOMPLETE SOLVENT APPLICATION ERROR.
Description of Event or Problem
IT WAS REPORTED THAT A LIQUID LEAK OCCURRED DURING PRIMING. THIS OCCURRED BEFORE PATIENT USE/DURING PRIMING AT FACILITY. THERE WAS NO PATIENT INVOLVEMENT, AND NO PATIENT HARM/ADVERSE EVENT REPORTED.