NRFIT EXTENSION SET W/ FILTER
Received Apr 17, 2025 · Event occurred Mar 18, 2025
Report 3012307300-2025-04404 · MDR key 21853092
Device
Generic name
Administrations Sets With Neuraxial Connectors
Manufacturer
Icu Medical, Inc., Minneapolis, MnCatalog number
21-7608-24
Lot number
UNKNOWN
Product problems
- Fluid/Blood Leak
- Fluid/Blood Leak
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.
Description of Event or Problem
IT WAS REPORTED THAT A CHEMICAL LEAK OCCURRED AT THE BASE OF THE FILTER AND LINE CONNECTION ON THE PUMP SIDE LINE. THE EVENT OCCURRED DURING PATIENT USE/ADMINISTRATION AT FACILITY. THERE WAS NO REPORTED PATIENT HARM/ADVERSE EVENT.
Additional Manufacturer Narrative
D3 - MFG ESTABLISHMENT NAME-CORRECTED. ONE DEVICE WAS RETURNED FOR ANALYSIS. A FUNCTIONAL TEST WAS PERFORMED AND THE REPORTED ISSUE WAS CONFIRMED. THE PROBABLE CAUSE OF THE REPORTED ISSUE WAS DUE TO THE LOOSE TUBING CONNECTION. A DEVICE HISTORY RECORD (DHR) REVIEW WAS CONDUCTED WHICH INDICATED ALL INSPECTIONS WERE COMPLETED AND NO ISSUES WERE NOTED DURING MANUFACTURE.