inforMED
MalfunctionFLL

LEVEL 1 ESOPHAGEAL / RECTAL 9F

Received Apr 17, 2025 · Event occurred Apr 9, 2025

Report 21852897 · MDR key 21852897

Device

Generic name

Thermometer, Electronic, Clinical

Model number

ER400-9

Catalog number

ER400-9

Product problems

  • Incorrect, Inadequate or Imprecise Result or Readings

Patient

5 MO · Male

  • Arrhythmia
  • Fever
  • Arrhythmia
  • Fever

Narrative

Description of Event or Problem

A RECTAL TEMP PROBE HAD BEEN PLACED FOR PATIENT FROM THE OR. THE NURSE (RN) CHECKED IT, PUSHED IT IN FURTHER, WAS READING 34-36C. THROUGH THE NIGHT, THE PATIENT HAD INCREASING ARRHYTHMIA. TEMP CHECKS ON THE RECTAL PROBE CONTINUED TO READ AFEBRILE DESPITE GOOD POSITIONING. RN REPLACED PROBE AND TEMP READ 38.2C. ESOPHAGEAL PROBE PLACED AND READ 40.2C. AN UNDETECTED FEVER CONTRIBUTED TO A BAD ARRHYTHMIA THAT COULD HAVE CAUSED HEMODYNAMIC INSTABILITY IN A TENUOUS CARDIAC POST OP PATIENT. ORIGINAL TEMP PROBE WAS FAULTY.