inforMED
InjuryOWB

INNOVA 4100

Received Apr 14, 2025 · Event occurred Jan 28, 2025

Report 9611343-2025-00006 · MDR key 21832426

Device

Generic name

Interventional Fluoroscopic X-ray

Product problems

  • Adverse Event Without Identified Device or Use Problem
  • Adverse Event Without Identified Device or Use Problem

Patient

58 YR · Male

  • Radiation Burn
  • Radiation Burn

Narrative

Description of Event or Problem

ON MARCH 21, 2025, A GE HEALTHCARE (GEHC) TECHNICAL SUPPORT ENGINEER WAS INFORMED BY A CUSTOMER THAT A PATIENT SUSTAINED A RADIATION INJURY TO THE SKIN FOLLOWING A LENGTHY PROCEDURE WITH A GE HEALTHCARE INNOVA 4100 SYSTEM ON (B)(6), 2025. AN INVESTIGATION IS IN PROGRESS.

Additional Manufacturer Narrative

LEGAL MANUFACTURER: HCS BUC - 283 RUE DE LA MINIERE FRANCE BUC YVELINES, 78530. THE GE HEALTHCARE DEVICE HAS NO UDI SINCE IT WAS MANUFACTURED BEFORE UDI IMPLEMENTATION.

Additional Manufacturer Narrative

GE HEALTHCARE HAS COMPLETED ITS INVESTIGATION. A REVIEW OF THE SYSTEM LOGS CONFIRMED THAT THE SYSTEM OPERATED WITHIN ITS SPECIFICATIONS. THE ADDITIONAL X-RAY DOSE WAS ATTRIBUTED TO THE OPERATOR DECISION. THE PATIENTÂS INJURY WAS DETERMINED TO BE AN ISOLATED CASE, RESULTING FROM A COMPLEX AND PROLONGED FLUOROSCOPIC PROCEDURE. FOR LARGER PATIENTS, SUCH PROCEDURES MAY LEAD TO SKIN INJURIES WHICH IS A KNOWN SIDE EFFECT. IT IS THE PHYSICIAN'S RESPONSIBILITY TO ASSESS THE BENEFITS VERSUS THE POTENTIAL RISKS FOR EACH PATIENT. THE INVESTIGATION IS NOW CLOSED, AND NO FURTHER ACTIONS ARE PLANNED BY GE HEALTHCARE.