AMSCO
Received Apr 11, 2025 · Event occurred Mar 15, 2025
Report 3005899764-2025-00031 · MDR key 21828979
Device
Generic name
Sterilizer
Manufacturer
Steris Mexico, S. De R.l. De C.v.Product problems
- Leak/Splash
- Leak/Splash
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
A STERIS TECHNICIAN ARRIVED ONSITE TO INSPECT THE UNIT AND FOUND THAT THE SOLENOID VALVE THAT FEEDS HOT WATER TO THE STEAM GENERATOR FAILED WHICH CAUSED THE WATER LEAK OF THE REPORTED EVENT. TO RESOLVE THE ISSUE, THE TECHNICIAN REPLACED THE SOLENOID VALVE, TESTED THE UNIT, FOUND IT TO BE OPERATING ACCORDING TO SPECIFICATION, AND RETURNED IT BACK TO SERVICE. NO ADDITIONAL ISSUES HAVE BEEN REPORTED. UDI RELATED DATA CLARIFICATION - THE DEVICE SUBJECT OF THE EVENT WAS MANUFACTURED PRIOR TO UDI COMPLIANCE, THEREFORE, UDI IS NOT APPLICABLE AND WAS NOT INCLUDED IN THE MDR.
Description of Event or Problem
THE USER FACILITY REPORTED THAT THEIR AMSCO 400 STERILIZER LEAKING WATER ONTO THE FLOOR. NO REPORT OF INJURY.