BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE
Received Apr 10, 2025 · Event occurred Mar 13, 2025
Report 1917413-2025-00439 · MDR key 21817424
Device
Generic name
Lymphocyte Separation Medium
Manufacturer
Becton, Dickinson & Co. (broken Bow)Catalog number
362761
Lot number
UNKNOWN
Product problems
- Incorrect, Inadequate or Imprecise Result or Readings
- Incorrect, Inadequate or Imprecise Result or Readings
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
IT WAS REPORTED DURING USE OF BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE, AN UNSPECIFIED NUMBER OF SAMPLES EXHIBITED POOR VIABILITY OF PERIPHERAL BLOOD MONONUCLEAR CELLS (PBMC'S). THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.
Additional Manufacturer Narrative
D4. MEDICAL DEVICE EXPIRATION DATE: UNKNOWN. H4. DEVICE MANUFACTURE DATE: UNKNOWN. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
Additional Manufacturer Narrative
INVESTIGATION SUMMARY: BD HAD NOT RECEIVED ANY SAMPLES FOR INVESTIGATION. THE DEVICE HISTORY RECORDS COULD NOT BE REVIEWED AS THE LOT NUMBER IS UNKNOWN. FURTHER TESTING COULD NOT BE PERFORMED AS THE LOT NUMBER IS UNKNOWN. THIS COMPLAINT HAS NOT BEEN CONFIRMED FOR THE INDICATED FAILURE MODE: ERRONEOUS RESULTS (POOR VIABILITY). NO DEFINITIVE ROOT CAUSE COULD BE ESTABLISHED FOR THE REPORTED DEFECT. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR THE IDENTIFICATION OF EMERGING TRENDS.
Description of Event or Problem
IT WAS REPORTED DURING USE OF BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM CITRATE, AN UNSPECIFIED NUMBER OF SAMPLES EXHIBITED POOR VIABILITY OF PERIPHERAL BLOOD MONONUCLEAR CELLS (PBMC'S). THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.