DJO SURGICAL
Received Apr 4, 2025 · Event occurred Dec 31, 2024
Report 1644408-2025-00386 · MDR key 21780956
Device
Generic name
Drill Bit, 3.2x35mm, Bayonet
Manufacturer
Encore Medical L.p.Catalog number
803-15-008
Lot number
325184L02
Product problems
- Mechanical Problem
- Detachment of Device or Device Component
- Mechanical Problem
- Detachment of Device or Device Component
Patient
64 YR · Unknown
- Foreign Body In Patient
- No Clinical Signs, Symptoms or Conditions
- Foreign Body In Patient
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
INSTRUMENT FAILURE - AT TIME OF SURGICAL PREPARATION FOR CANCELLOUS BONE SCREW DURING A TOTAL HIP ARTHROPLASTY PROCEDURE, DURING NORMAL USE OF THE DRILL, THE BIT BROKE OFF WHILE INSIDE THE PATIENT. PER SURGEON, MULTIPLE ATTEMPTS TO REMOVE THE BROKEN DRILL BIT WERE UNSUCCESSFUL. HE JUDGED FURTHER ATTEMPTS TO REMOVE THE RETAINED RILL BIT AS HAVING THE POTENTIAL FOR SIGNIFICANT DAMAGE TO THE BONE AND DECIDED TO PROCEED WITH THE SURGERY, LEAVING THE RETAINED DRILL BIT IN PLACE. THE RECOVERED PORTION OF THE DRILL BIT WAS MEASURED AT 15MM. SURGEON THEN ESTIMATED THE LENGTH OF THE RETAINED PORTION OF THE DRILL BIT AT 20MM. SURGEON ANTICIPATES NO EFFECTS TO PATIENT OF RETAINED PIECE.
Additional Manufacturer Narrative
ADDITIONAL REPORTING ON THIS EVENT WILL BE PROVIDED AS A SUPPLEMENTAL REPORT TO THIS DOCUMENT AS SOON AS IT BECOMES AVAILABLE
Additional Manufacturer Narrative
SEE H4, H10, AND H11. COMPLAINT HAS BEEN EVALUATED AND IS SIMILAR TO REPORT NUMBER 1644408-2024-02011; 803-15-006, S303 - BROKE/CRACKED/DAMAGED, INSTRUMENT FAILURE. IF ADDITIONAL INFORMATION REGARDING THE REPORTED EVENT IS SUBMITTED AT A FUTURE DATE, THIS INVESTIGATION WILL BE RE-EVALUATED.