WARMER,HEEL,INFANT,MTL DISC,CLOTH STRAP
Received Apr 4, 2025 · Event occurred Feb 15, 2025
Report 1417592-2025-00162 · MDR key 21774883
Device
Generic name
Warmer,heel,infant,mtl Disc,cloth Strap
Manufacturer
Medline Industries, LpCatalog number
MDS138009
Lot number
VN24221A27
Product problems
- Temperature Problem
- Temperature Problem
Patient
NA · Unknown
- Burn(s)
- Burn(s)
Narrative
Additional Manufacturer Narrative
ACCORDING TO THE CUSTOMER, ON 02/15/2025 THERE WERE "SEVERAL HEEL WARMERS THAT HAD BEEN ATTEMPTED ON THE PATIENT AND LEFT IN THE INCUBATOR", "THE INFANT'S ARM WAS LAYING ON THE WARMERS", AND IT CAUSED A "SEVERE BURN" TO THE INFANTS "MID ARM / ELBOW". THE CUSTOMER REPORTED "THE INCUBATOR IS UNDER A HEAT LAMP AND THE WARMERS POTENTIALLY REHEATED". THE CUSTOMER REPORTED "TREATMENT FOR THE BURN WAS ADMINISTERED" BUT THE SPECIFIC CARE ADMINISTERED COULD NOT BE PROVIDED. THE CUSTOMER REPORTED THE INFANT "MAY REQUIRE PLASTIC SURGERY TO REPAIR THE BURN". NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. IT HAS BEEN DETERMINED THAT THE REPORTED EVENT CAUSED OR CONTRIBUTED TO SERIOUS INJURY, THEREFORE, THIS MEDWATCH IS BEING FILED. IF ANY FURTHER RELEVANT INFORMATION IS IDENTIFIED OR OBTAINED, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED.
Description of Event or Problem
ACCORDING TO THE CUSTOMER, ON 02/15/2025 THERE WERE "SEVERAL HEEL WARMERS THAT HAD BEEN ATTEMPTED ON THE PATIENT AND LEFT IN THE INCUBATOR", "THE INFANT'S ARM WAS LAYING ON THE WARMERS", AND IT CAUSED A "SEVERE BURN" TO THE INFANTS "MID ARM / ELBOW."