INTERLOCKING BOLT FOR 4.5MM VA-LCP CONDYLAR INSERTN HANDLE
Received Apr 2, 2025 · Event occurred Jan 1, 2025
Report 8030965-2025-02827 · MDR key 21754504
Device
Generic name
Condylar Plate Fixation Implant
Manufacturer
Synthes GmbhCatalog number
03.231.005
Lot number
2L08427
Product problems
- Device-Device Incompatibility
- Material Twisted/Bent
- Device-Device Incompatibility
- Material Twisted/Bent
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. H11 ADDITIONAL NARRATIVE: ADDED: D10. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.
Additional Manufacturer Narrative
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. H11 ADDITIONAL NARRATIVE: ADDED: D4 (EXP. DATE, LOT), D9, H4. CORRECTED: D4 (PRIMARY UDI NUMBER). H3, H6: THE DEVICE WAS RECEIVED, THE INVESTIGATION IS IN PROGRESS, NO CONCLUSION COULD BE DRAWN AT THE TIME OF FILING THIS REPORT. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
Additional Manufacturer Narrative
PRODUCT COMPLAINT # (B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803 (AND/OR PART 4, AS APPLICABLE). THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. H11 ADDITIONAL NARRATIVE: D4: UDI: AS THE LOT NUMBER FOR THE DEVICE INVOLVED IN THE EVENT WAS NOT PROVIDED, THE FULL UDI IS CURRENTLY NOT AVAILABLE.
Description of Event or Problem
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE YELLOW STRIPED CONNECTING SCREW DID NOT THREAD TIGHTLY ONTO THE VA CONDYLAR PLATE. IT WAS THOUGHT THAT THE SCREW MUST BE STRIPPED. IT WOULDN'T THREAD INTO ANY PLATE. THE ISSUE HAD OCCURRED WITH THE SURGEON THREE TIMES. DURING SURGERY ON (B)(6) 2025, THERE WAS A 45-MINUTE SURGICAL DELAY IN RELATION TO THE EVENT. THERE WERE NO PATIENT CONSEQUENCES.
Additional Manufacturer Narrative
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. H11 ADDITIONAL NARRATIVE: CORRECTED: D4 (UDI), G1 AND H4. H3, H6: THE PRODUCT WAS RETURNED TO J&J MEDTECH ORTHOPAEDICS FOR EVALUATION. VISUAL INSPECTION OF THE RETURNED DEVICE FOUND THE THREADS BOTH AT THE TIP AND SHAFT WERE STRIPPED. THE OBSERVED CONDITION WAS IDENTIFIED AS AN END OF LIFE INDICATOR; DAMAGE CONSISTENT WITH REPEATED USE AND SERVICING. THE LIFECYCLE REQUIREMENTS OF THE DEVICE ARE EVENT RELATED AND DEPEND ON THE USE AND INSPECTION OF THE DEVICE IN CLINICAL PRACTICE. AS THE DEVICE CAN BE DAMAGED ON THE FIRST OR 100TH USE, THE DEVICE MUST BE PROPERLY INSPECTED PRIOR TO EACH SURGICAL USE. REFER TO THE DEVICE/COUNTRY SPECIFIC IFU FOR INFORMATION RELATED TO END OF LIFE, REPROCESSING INSTRUCTIONS, AND INSPECTION PROCEDURES. A DIMENSIONAL INSPECTION WAS NOT PERFORMED DUE TO POST MANUFACTURING DAMAGE. A FUNCTIONAL EVALUATION WAS NOT PERFORMED DUE TO POST MANUFACTURING DAMAGE. HOWEVER, BASED ON THE OBSERVED DAMAGE, IT IS NOT UNREASONABLE TO CONCLUDE THE DEVICE WOULD PRESENT ASSEMBLY ISSUES. THE ALLEGATION OF UNABLE TO ASSEMBLE CAN BE CONFIRMED. THE OVERALL COMPLAINT WAS CONFIRMED AS THE OBSERVED CONDITION OF THE 03.231.005 WOULD HAVE CONTRIBUTED TO THE COMPLAINED DEVICE ISSUE. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE REPORTED COMPLAINT CONDITION. AS PART OF J&J MEDTECH ORTHOPAEDICS QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE HISTORY REVIEW: PART# 03.231.005. LOT # 2L08427. MANUFACTURING SITE: WERK HÄGENDORF. SUPPLIER : NA. RELEASE TO WAREHOUSE DATE:04 DEC 2018. EXPIRATION DATE: NA. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED ARTICLE LOT AND NO NON-CONFORMANCES WERE IDENTIFIED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.