MalfunctionCKA

ALANINE AMINOTRANSFERASE ACC. TO IFCC

Received Apr 2, 2025 · Event occurred Mar 19, 2025

Report 1823260-2025-00984 · MDR key 21752473

Device

Generic name

Nadh Oxidation/nad Reduction, Alt/sgpt

Manufacturer

Roche Diagnostics

Catalog number

05850797190

Lot number

81382001

Product problems

  • Incorrect Measurement
  • Incorrect Measurement

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE COBAS 8000 COBAS C 701 MODULE SERIAL NUMBER WAS (B)(6). THE AST REAGENT USED ON THE COBAS PRO C503 ANALYZER WAS EXPIRED AT THE TIME OF THE SAMPLE TESTING. THE INVESTIGATION DID NOT IDENTIFY A PRODUCT PROBLEM. THE CAUSE OF THE EVENT COULD NOT BE DETERMINED.

Description of Event or Problem

THERE WAS AN ALLEGATION OF QUESTIONABLE ALANINE AMINOTRANSFERASE ACC. TO IFCC (ALT) RESULTS FROM THE COBAS 8000 COBAS C 701 MODULE. THE CUSTOMER SUSPECTED AN INTERFERENCE. THE INITIAL RESULT WAS 0.2 U/L. THE REPEAT RESULT WITH A DECREASED SAMPLE VOLUME WAS -3.7 U/L WITH A DATA FLAG. THE REPEAT RESULT WITH AN INCREASED SAMPLE VOLUME WAS 1.4 U/L. THE RESULT USING THE J&J (DRY CHEMICAL TEST) METHOD WAS 43.8 U/L. THE CUSTOMER TESTED THE SAMPLE USING THE ALTP GEN.2 REAGENT ON A COBAS PRO C503 ANALYZER AND THE RESULT WAS 71.8 U/L THE QUESTIONABLE RESULTS WERE NOT REPORTED OUTSIDE OF THE LABORATORY.