inforMED
MalfunctionIOQ

ENVISION BLOWER KIT

Received Apr 1, 2025 · Event occurred Mar 21, 2025

Report 1824206-2025-00602 · MDR key 21740818

Device

Generic name

Bed, Flotation Therapy, Powered

Model number

P136026

Catalog number

P136026

Product problems

  • Material Split, Cut or Torn
  • Material Split, Cut or Torn

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

D4: UNIQUE IDENTIFIER (UDI) #: THE MANUFACTURING DATE (11) IS UNKNOWN AT THIS TIME; THEREFORE, THE UDI NUMBER ONLY WILL CONTAIN SERIAL NUMBER (21). H11: THE BAXTER TECHNICIAN FOUND THE POWER CORD NEEDED TO BE REPLACED. PER THE BAXTER SERVICE MANUAL, EXAMINE THE POWER CORD AND ITS PLUG FOR DAMAGE. EXAMINE THE LENGTH OF THE POWER CORD FOR CUTS AND ABRASIONS. PLUG THE POWER CORD INTO A CALIBRATED SAFETY ANALYZER AND ATTACH THE ANALYZER TO THE EXPOSED METAL CHASSIS GROUND TEST POINT TO MAKE SURE THE GROUND RESISTANCE IS NO MORE THAN 200 MILLIOHMS. PLUG THE POWER CORD INTO A CALIBRATED SAFETY ANALYZER AND MAKE SURE THE LEAKAGE CURRENT IS NO MORE THAN 300 MICROAMPS IN NORMAL CONDITION (NC). REPLACE AS NECESSARY. A SEARCH OF THE BAXTER MAINTENANCE RECORDS SHOWED BAXTER PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN AUG 13, 2024. IT IS UNKNOWN IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECHNICIAN REPLACED THE POWER CORD TO RESOLVE THE REPORTED EVENT. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem

IT WAS REPORTED ENVISION BLOWER KIT (PRODUCT CODE P136026, SERIAL NUMBER (B)(6)), HAD POWER CORD DAMAGE WITH EXPOSED COPPER WIRES. THIS OCCURRED DURING BAXTER SERVICING/TESTING. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.