MalfunctionDQA
SPO2 CABLE
Received Apr 1, 2025 · Event occurred Mar 14, 2025
Report 21738453 · MDR key 21738453
Device
Generic name
Oximeter
Manufacturer
Philips Medizin Systeme Boeblingen GmbhModel number
M1943ALCatalog number
989803128651
Lot number
1224
Patient
65 YR · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
PHILIPS PULSE OXIMETER ADAPTER CABLE ATTACHED TO SENSOR/FINGER PROBE THAT HAD BEEN APPLIED TO THE PATIENT AND CONNECTED TO THE MACHINE. MACHINE NOTED "SENSOR NOT FOUND." NEW FINGER PROBE APPLIED AND RECEIVED SAME ERROR MESSAGE. CABLE EXCHANGED FOR NEW CABLE AND MACHINE REGISTERED SPO2 [SATURATION OF PERIPHERAL OXYGEN] READING. THIS RESULTED IN A DELAY IN PATIENT CARE, NO KNOWN HARM TO THE PATIENT AT THIS TIME.