inforMED
MalfunctionJCF

BD VACUTAINER® CPT¿ NH: ~130 IU FICOLL¿: 2.0ML

Received Mar 28, 2025 · Event occurred Jul 31, 2024

Report 1917413-2025-00344 · MDR key 21726742

Device

Generic name

Lymphocyte Separation Medium

Catalog number

362780

Lot number

3354407

Product problems

  • Incorrect, Inadequate or Imprecise Result or Readings
  • Incorrect, Inadequate or Imprecise Result or Readings

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

INVESTIGATION SUMMARY: BD RECEIVED 2 PHOTOS FOR INVESTIGATION. THE PHOTOS WERE REVIEWED AND THE INDICATED FAILURE MODE FOR POOR BARRIER SEPARATION WAS OBSERVED. ADDITIONALLY, 44 RETENTION SAMPLES FROM BD INVENTORY WERE VISUALLY INSPECTED WITH NO ISSUES OBSERVED. COMPLAINTS FOR SAMPLE QUALITY ARE UNDER STATISTICAL CONTROL FOR THE MONTH OF MARCH 2025. AT THIS TIME, FURTHER TESTING IS NOT INDICATED. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT HAS BEEN CONFIRMED FOR THE INDICATED FAILURE MODE POOR BARRIER SEPARATION BASED ON THE PHOTOS PROVIDED. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS. BD QUALITY WILL CONTINUE TO MONITOR SAMPLE QUALITY COMPLAINTS.

Description of Event or Problem

IT WAS REPORTED WHILE USING BD VACUTAINER® CPT¿ NH: 130 IU FICOLL¿: 2.0ML, AN UNSPECIFIED NUMBER OF SAMPLES EXHIBIT POOR BARRIER SEPARATION UPON PROCESSING. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

Additional Manufacturer Narrative

E1. INITIAL REPORTER FACILITY NAME: (B)(6). A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem