inforMED
MalfunctionCCC

RAPIDPOINT 400

Received Mar 30, 1999 · Event occurred Mar 1, 1999

Report 1217157-1999-00001 · MDR key 217100

Device

Generic name

Blood Gas Analyzer

Model number

NA

Catalog number

109824

Lot number

NA

Product problems

  • Other (for use when an appropriate device code cannot be identified)

Patient

NA

Narrative

Description of Event or Problem

CUSTOMER REPORTED UNDER-RECOVERY OF RAPIDQC COMPLETE QUALITY CONTROL MATERIAL FOR PARTIAL OXYGEN PRESSURE RESULTS. LEVEL 1 : ASSIGNED RANGE 134-156 MMHG; VALUES OBTAINED 126/127/130 MMHG; LEVEL 2: ASSIGNED RANGE 88-104; VALUES OBTAINED 83/85/97; LEVEL 3: ASSIGNED RANGE 12-34; VALUES OBTAINED 21/22. THIS INDICATED THAT THERE WAS POTENTIAL FOR UNDER-REPORTING PT VALUES FOR PARTIAL PRESSURE OF OXYGEN. IT IS BELIEVED THAT NO PT DATA WAS REPORTED.

Remedial action

  • Notification
  • Relabeling