CADD SOLIS COMM MODULE
Received Mar 27, 2025 · Event occurred May 1, 2024
Report 3012307300-2025-03328 · MDR key 21709156
Device
Generic name
Accessories, Pump, Infusion
Manufacturer
Smiths Medical Asd, Inc.Catalog number
21-2131-25
Lot number
18489537
Product problems
- Wireless Communication Problem
- Wireless Communication Problem
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
ONE DEVICE WAS RETURNED FOR ANALYSIS. FUNCTIONAL AND VISUAL TESTS WERE PERFORMED, AND THE REPORTED ISSUE WAS NOT DUPLICATED. THE CAUSE OF THE REPORTED ISSUE WAS DUE TO A MALFUNCTIONING WIRELESS RADIO BOARD. A DHR REVIEW IS NOT APPLICABLE IN THIS CASE BECAUSE ITEM IS A PURCHASED FINISHED GOOD, MANUFACTURING RECORDS ARE RETAINED BY THE SUPPLIER AT THEIR MANUFACTURING SITE AND ARE NOT READILY AVAILABLE FOR REVIEW.
Description of Event or Problem
IT WAS REPORTED THAT THE PUMP'S COMM-MODULE WAS DEFECTIVE, AND THE WIRELESS MODULE WAS EXPERIENCING AN INTERMITTENT CONNECTION WITH THE PUMP. THERE WAS NO PATIENT INVOLVEMENT, AND NO ADVERSE EVENT WAS REPORTED.