TOTALCARE BARIATRIC PLUS BED

Received Mar 26, 2025 · Event occurred Mar 16, 2025

Report 1824206-2025-00561 · MDR key 21699626

Device

Generic name

Bed, Flotation Therapy, Powered

Model number

P1840D

Catalog number

P1840DRE0001

Product problems

  • Unintended Movement
  • Unintended Movement

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THAT TOTALCARE BARIATRIC RENTAL-NEW (PRODUCT CODE P1840DRE0001, SERIAL NUMBER (B)(6)), HAD BRAKES NOT HOLDING. THIS OCCURRED DURING BAXTER SERVICING/TESTING. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.

Additional Manufacturer Narrative

THE BAXTER TECHNICIAN FOUND THE CASTERS NEEDED TO BE REPLACED. PER THE BAXTER SERVICE MANUAL, THE TOTALCARE BARIATRIC BED AND TOTALCARE BARIATRIC PLUS THERAPY SYSTEM REQUIRE AN EFFECTIVE MAINTENANCE PROGRAM. WE RECOMMEND THAT YOU PERFORM SEMIANNUAL PREVENTIVE MAINTENANCE (PM). PM WILL MINIMIZE DOWNTIME DUE TO EXCESSIVE WEAR. INSPECT THE CASTERS FOR PROPER MOUNTING, EXCESSIVE WEAR, OR TREAD DAMAGE. REPLACE AS NEEDED. VERIFY PROPER OPERATION OF THE BRAKE SYSTEM. ADJUST AS NECESSARY. A SEARCH OF THE BAXTER MAINTENANCE RECORDS SHOWED BAXTER PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN (B)(6) 2025. IT IS UNKNOWN IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECHNICIAN REPLACED THE CASTERS TO RESOLVE THE REPORTED EVENT. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.