TOTALCARE BARIATRIC PLUS BED
Received Mar 25, 2025 · Event occurred Mar 11, 2025
Report 1824206-2025-00551 · MDR key 21690548
Device
Generic name
Bed, Flotation Therapy, Powered
Manufacturer
Baxter Healthcare CorporationModel number
P1840DCatalog number
P1840DRE0001
Lot number
NI
Product problems
- Unintended Movement
- Unintended Movement
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN TO REPORT THE TOTAL CARE BARIATRIC RENTAL-NEW BRAKE CASTERS WERE NOT HOLDING. THE BED WAS LOCATED AT THE ACCOUNT. HIS OCCURRED DURING BAXTER SERVICING/TESTING. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED. THERE WERE NO OTHER DEVICES REPORTED TO BE INVOLVED. THE CUSTOMER DID NOT COMMUNICATE THIS EVENT TO ANOTHER BAXTER DEPARTMENT OR AUTHORITY.
Additional Manufacturer Narrative
THE BAXTER TECHNICIAN FOUND THE CASTERS NEEDED TO BE REPLACED. PER THE BAXTER SERVICE MANUAL, THE TOTAL CARE BARIATRIC BED AND TOTAL CARE BARIATRIC PLUS THERAPY SYSTEM REQUIRE AN EFFECTIVE MAINTENANCE PROGRAM. WE RECOMMEND THAT YOU PERFORM SEMIANNUAL PREVENTIVE MAINTENANCE (PM). PM WILL MINIMIZE DOWNTIME DUE TO EXCESSIVE WEAR. INSPECT THE CASTERS FOR PROPER MOUNTING, EXCESSIVE WEAR, OR TREAD DAMAGE. REPLACE AS NEEDED. VERIFY PROPER OPERATION OF THE BRAKE SYSTEM. ADJUST AS NECESSARY. A SEARCH OF THE BAXTER MAINTENANCE RECORDS DID NOT SHOW BAXTER PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNKNOWN IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR BEDS. THE TECHNICIAN REPLACED THE CASTERS TO RESOLVE THE REPORTED EVENT. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.