Received Mar 21, 2025 · Event occurred Feb 20, 2025

Report 2024168-2025-02985 · MDR key 21665758

Device

Generic name

Adaptor, Stopcock, Manifold, Fitting, Cardiopulmonary Bypass

Model number

1003331

Catalog number

1003331

Lot number

60573590

Product problems

  • Leak/Splash
  • Leak/Splash

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE DEVICE WAS NOT RETURNED FOR ANALYSIS. PRODUCTION RECORD AND CORRECTIVE AND PREVENTATIVE ACTIONS (CAPA) REVIEWS WERE PERFORMED AND REVEALED NO INDICATION OF A PRODUCT QUALITY ISSUE. ADDITIONALLY, A QUERY OF THE COMPLAINT HANDLING DATABASE FOR THE REPORTED LOT REVEALED THERE IS NO INDICATION OF A LOT SPECIFIC ISSUE. IN THIS CASE, IT IS POSSIBLE THAT THE CAP SEAL WAS NOT FULLY TIGHTENED, THEREFORE RESULTING IN THE REPORTED LEAK DIFFICULTIES. HOWEVER, THE DEVICE WAS NOT RETURNED FOR ANALYSIS; THEREFORE, IT CANNOT BE CONFIRMED. THE INVESTIGATION DETERMINED A CONCLUSIVE CAUSE FOR THE REPORTED DIFFICULTY CANNOT BE DETERMINED. BASED ON THE RESULTS OF THE COMPLAINT INVESTIGATION, THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO THE DESIGN, MANUFACTURE, OR LABELING OF THE DEVICE.

Description of Event or Problem

IT WAS REPORTED THAT THE COPILOT WAS LEAKING BLOOD FROM THE BLEEDBACK CONTROL VALVE DURING USE. THE DEVICE WAS CONNECTED TO A CATHETER, BALLOON AND WIRE. THE SAME DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.