MalfunctionJAQ

GAMMAMED PLUS

Received Jul 5, 2011 · Event occurred Jun 5, 2011

Report 9612638-2011-00005 · MDR key 2166279

Device

Generic name

Sys, Applicator, Radionuclide, R-c

Model number

H64

Product problems

  • Loose or Intermittent Connection

Patient

Not reported

  • No Known Impact Or Consequence To Patient

Narrative

Description of Event or Problem

TEMPLATE REPORTED AS NOT CLAMPING NEEDLES SECURELY. NEEDLES ARE NOT HELD FAST TO THE TEMPLATE LOCKING MECHANISM. A USER REPORTS THAT INITIALLY THEY WERE LOCKING SUFFICIENTLY, BUT AFTER HANDLING ACCORDING TO DOCUMENTATION, TWO SUBSEQUENT IMPLANTS WERE INTERRUPTED FOR RE-SCANNING BECAUSE THE NEEDLES HAD SHIFTED.

Additional Manufacturer Narrative

THERE WAS NO INJURY ASSOCIATED WITH THIS COMPLAINT. USER HAD TO ADJUST NEEDLES AND RE-SCAN RESULTING IN A MINOR TREATMENT DELAY. LARGE SHIFTS IN THE NEEDLE TEMPLATE POSITIONS WOULD BE OBVIOUS TO THE USER PRIOR TO TREATMENT. METHOD - THOUGH STILL UNDER INVESTIGATION, VARIAN HAS DETERMINED THAT A MDR IS APPROPRIATE AS THIS REPORTED MALFUNCTION, SHOULD IT RECUR, MAY POTENTIALLY RESULT IN SERIOUS INJURY. ADDITIONAL FOLLOW-UP TO THIS MDR IS EXPECTED UPON COMPLETION OF THE INVESTIGATION.