ELECSYS VITAMIN B12 II
Received Mar 21, 2025 · Event occurred Feb 4, 2025
Report 1823260-2025-00821 · MDR key 21662738
Device
Generic name
Radioassay, Vitamin B12
Manufacturer
Roche DiagnosticsCatalog number
07028121190
Lot number
808573
Product problems
- Low Readings
- Non Reproducible Results
- Low Readings
- Non Reproducible Results
Patient
73 YR · Female
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THE ISSUE WAS DETECTABLE TO THE USER AS THE QC WAS CONSISTENTLY LOW OUT OF RANGE. BASED ON THE PROVIDED INFORMATION, THE ROOT CAUSE WAS CONSISTENT WITH AN ONBOARD STABILITY ISSUE OF VITAMIN B12 II KITS ON THE COBAS PURE E402 ANALYTICAL UNIT. A QUALITY NOTIFICATION WAS PUBLISHED.
Additional Manufacturer Narrative
THE COBAS PURE E402 ANALYTICAL UNIT SERIAL NUMBER WAS (B)(6). THE QC TRACE SHOWED CAL.E ALARM ON 11-JAN-2025 SUGGESTING A CALIBRATION ISSUE. THE QC FROM 03-FEB-2025 AND 04-FEB-2025 WAS CONSISTENTLY BORDERLINE OR OUT OF RANGE. NO QC DATA WAS PROVIDED FROM 05-FEB-2025. THE INVESTIGATION IS ONGOING.
Description of Event or Problem
WE RECEIVED AN ALLEGATION ABOUT DISCREPANT RESULTS FOR 4 PATIENTS' SAMPLES TESTED WITH ELECSYS VITAMIN B12 G2 (VITAMIN B12 II) ASSAY ON A COBAS PURE E402 ANALYTICAL UNIT. SAMPLE 1: INITIAL RESULT: 350 PG/ML. REPEAT RESULT: 669 PG/ML. SAMPLE 2: INITIAL RESULT: 100 PG/ML (ACCOMPANIED BY A DATA FLAG). REPEAT RESULT: 237 PG/ML. SAMPLE 3 AND SAMPLE 4 WERE TESTED ON (B)(6)2025: SAMPLE 3: INITIAL RESULT: 214 PG/ML. REPEAT RESULT: 375 PG/ML. SAMPLE 4: INITIAL RESULT: 224 PG/ML. REPEAT RESULT: 419 PG/ML. THE CUSTOMER HAD CALIBRATION AND QC ISSUES PROMPTING THE REPEAT OF THE SAMPLES IN ANOTHER LABORATORY. THE REPEAT RESULTS WERE DEEMED TO BE CORRECT.