CADD SOLIS COMM MODULE
Received Mar 20, 2025 · Event occurred Oct 1, 2024
Report 3012307300-2025-03117 · MDR key 21657349
Device
Generic name
Accessories, Pump, Infusion
Manufacturer
Smiths Medical Asd, IncCatalog number
21-2131-25
Lot number
18489537
Product problems
- Communication or Transmission Problem
- Communication or Transmission Problem
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
IT WAS REPORTED THAT DEVICE HAS WIRELESS MODULE INTERMITTENT CONNECTION ISSUES. THERE WAS UNKNOWN PATIENT INVOLVEMENT AND UNKNOWN PATIENT HARM/ADVERSE EVENT REPORTED.
Additional Manufacturer Narrative
ONE DEVICE WAS RETURNED FOR ANALYSIS IN GOOD CONDITION. A DHR REVIEW IS NOT APPLICABLE IN THIS CASE BECAUSE ITEM IS A PURCHASED FINISHED GOOD, MANUFACTURING RECORDS ARE RETAINED BY THE SUPPLIER AT THEIR MANUFACTURING SITE AND ARE NOT READILY AVAILABLE FOR REVIEW. EVALUATION FOUND THE COMM MODULE CONNECTING TO THE PUMP PROPERLY AND WAS BEING DETECTED WITHOUT ERRORS WHEN PAIRED WITH A KNOWN GOOD OPERATING PUMP. EVALUATION DID FIND THE COMM MODULE FAILING TO ENABLE ITS WIRELESS SETTINGS AND CONNECT TO THE WIRELESS NETWORK. THE MOST PROBABLE CAUSE IS A MALFUNCTIONING WIRELESS RADIO BOARD. DEVICE IS BEING SENT TO SUPPLIER FOR FURTHER ANALYSIS. B3. DATE OF EVENT: UNKNOWN. NO INFORMATION HAS BEEN PROVIDED TO DATE.