7.5CM LARGE CRANIOTOME ATTCH
Received Mar 20, 2025 · Event occurred Jan 1, 2025
Report 1045834-2025-00191 · MDR key 21653299
Device
Generic name
Drill, Surgical, Ent (electric Or Pneumatic) Including Handpiece
Manufacturer
Depuy Synthes Products LlcModel number
SYNTHES GMBHCatalog number
CRANI-L
Product problems
- Detachment of Device or Device Component
- Detachment of Device or Device Component
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THIS MEDWATCH, A SUPPLEMENTAL MEDWATCH REPORT WILL BE FILED AS APPROPRIATE. H11: ADDITIONAL NARRATIVE: THE ACTUAL DEVICE HAS BEEN RETURNED AND IS CURRENTLY PENDING EVALUATION. ONCE THE DEVICE HAS BEEN EVALUATED, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SENT ACCORDINGLY.
Description of Event or Problem
IT WAS REPORTED THAT THE CRANIOTOME DEVICE FELL APART. IT WAS NOT REPORTED IF THE DEVICE WAS USED IN SURGERY. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN A SURGICAL PROCEDURE OR IF A SPARE DEVICE WAS AVAILABLE. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2025. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Additional Manufacturer Narrative
THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THIS MEDWATCH, A SUPPLEMENTAL MEDWATCH REPORT WILL BE FILED AS APPROPRIATE. H11 ADDITIONAL NARRATIVE: CORRECTION: D9: THE DEVICE RETURNED DATE HAS BEEN CORRECTED. DEVICE EVALUATION: THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. DURING REPAIR, AN EVALUATION WAS PERFORMED, AND IT WAS DETERMINED THAT THE REPORTED CONDITION OF THE DEVICE FALLING APART WAS NOT CONFIRMED. THEREFORE, AN ASSIGNABLE ROOT CAUSE WAS NOT DETERMINED. HOWEVER, DURING EVALUATION, IT WAS DETERMINED THAT THE CRANIOTOME NEURO-TIP WAS BENT/DAMAGED AND THE DEVICE HAD COSMETIC DAMAGE AND MISSING COMPONENTS. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR VISUAL ASSESSMENT AND COLOR BAND VISUAL ASSESSMENT. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO COMPONENT FAILURE FROM NORMAL WEAR.