VITAL SIGNS
Received Mar 20, 2025 · Event occurred Mar 7, 2025
Report 21652088 · MDR key 21652088
Device
Generic name
Circuit, Breathing (w Connector, Adaptor, Y Piece)
Manufacturer
Sunmed Group Holdings, Llc Dba AirlifeModel number
AXXXXX61Catalog number
AXXXXX61
Lot number
0004304010
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
ANESTHESIA CIRCUIT HAD A KNOT IN THE INTERNAL CO2 DETECTION TUBING. MANUFACTURER RESPONSE FOR CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE), VITAL SIGNS¿ (PER SITE REPORTER). FROM SUNMED, "WE HAVE RECEIVED 1 OTHER COMPLAINT FOR THIS ISSUE, COMPLAINT-[REDACTED], FROM [REDACTED]. THOSE ARE THE ONLY 2 COMPLAINTS FOR THE ISSUE IN THE PAST 2 YEARS. THE KNOT IS NOT A COMMON OCCURRENCE. HOWEVER, WE ARE IMPLEMENTING CORRECTIVE ACTION TO AVOID THIS DEFECT. A FIXTURE IS BEING INSTALLED IN THE ASSEMBLY PROCESS TO DETECT ANY KNOT IN THE TUBING AND HALT MANUFACTURING. THE FIXTURE IS BEING IMPLEMENTED IN CHANGE REQUEST #[REDACTED]."