inforMED
MalfunctionCKA

SYNCHRON® CX TOTAL PROTEIN REAGENT

Received Jul 12, 2011 · Event occurred Jun 8, 2011

Report 2050012-2011-02543 · MDR key 2158881

Device

Generic name

Total Protein Test Reagent

Catalog number

442740

Lot number

Z102232

Product problems

  • Leak/Splash

Patient

Not reported

  • No Consequences Or Impact To Patient

Narrative

Description of Event or Problem

A CUSTOMER REPORTED TO BECKMAN COULTER, INC. OF A LEAK FROM A BOTTLE OF SYNCHRON CX TOTAL PROTEIN REAGENT. THERE WAS NO EXPOSURE TO UNCOVERED WOUNDS OR MUCOUS MEMBRANES. NO INJURY WAS REPORTED, AND MEDICAL ATTENTION WAS NOT SOUGHT. (B)(4) WAS REVIEWED AND THE FACILITY HAS EXPOSURE CONTROL PLAN. THERE WAS NO EFFECT TO PATIENT OR USER.

Additional Manufacturer Narrative

THE CRACK AT THE BOTTOM OF THE BOTTLE CAUSED THE LEAKAGE. PHOTOGRAPHS OF DAMAGED REAGENT KIT WERE RECEIVED. A REPLACEMENT KIT WAS SENT TO THE CUSTOMER.