inforMED
MalfunctionBSZ

ADU

Received Mar 11, 2025 · Event occurred Feb 11, 2025

Report 9610105-2025-00002 · MDR key 21572183

Device

Generic name

Anesthesia Gas Machine

Catalog number

NA

Lot number

NA

Product problems

  • Failure to Run on Battery
  • Failure to Run on Battery

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

A GE HEALTHCARE SERVICE REPRESENTATIVE PERFORMED A CHECKOUT OF THE SYSTEM AND CONFIRMED THE REPORTED ISSUE. THE BATTERY WAS REPLACED TO RESOLVE THE ISSUE. BLOCK A: PATIENT INFORMATION COULD NOT BE OBTAINED DUE TO COUNTRY PRIVACY LAWS. Â LEGAL MANUFACTURER: HCS HELSINKI - KUORTANEENKATU 2 FINLAND HELSINKI ETELA-SUOMEN LAANI, FIN-00510.

Description of Event or Problem

IT WAS REPORTED THAT THERE WAS A LOSS OF MECHANICAL VENTILATION DUE TO BATTERY FAILURE. THERE WAS NO PATIENT INJURY.