inforMED
MalfunctionPMT

VITROS IMMUNODIAGNOSTIC PRODUCTS BRAHMS PCT REAGENT PACK

Received Mar 10, 2025 · Event occurred Feb 11, 2025

Report 3007111389-2025-00020 · MDR key 21559949

Device

Generic name

In-vitro Diagnostics

Catalog number

6197735

Lot number

0440

Product problems

  • High Test Results
  • Non Reproducible Results
  • High Test Results
  • Non Reproducible Results

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE INVESTIGATION HAS DETERMINED THAT A NON-REPRODUCIBLE, HIGHER THAN EXPECTED VITROS PROCALCITONIN (PCT) RESULT WAS OBTAINED FROM A SAMPLE FROM A SINGLE PATIENT WHEN TESTED ON A VITROS XT 7600 INTEGRATED SYSTEM USING REAGENT LOT 0440 WHEN COMPARED TO A REPEAT PCT TEST RESULT OBTAINED FROM A REPEAT TEST OF THE SAME SAMPLE ON THE SAME VITROS XT 7600 INTEGRATED SYSTEM. A DEFINITIVE ASSIGNABLE CAUSE FOR THE NON-REPRODUCIBLE, HIGHER THAN EXPECTED VITROS PCT RESULT COULD NOT BE DETERMINED WITH THE INFORMATION PROVIDED. BASED ON HISTORICAL QUALITY CONTROL RESULTS, A VITROS PCT PERFORMANCE ISSUE IS NOT A LIKELY CONTRIBUTOR TO THE EVENT. IN ADDITION, THE REPEAT TEST RESULT OF 0.059 NG/ML WAS PROCESSED USING THE SAME VITROS PCT REAGENT PACK, FURTHER SUPPORTING THAT THE VITROS PCT REAGENT LOT 0440 IS NOT A CONTRIBUTOR TO THE EVENT. THERE IS ALSO NO INDICATION OF AN INSTRUMENT, HOWEVER, AN INSTRUMENT ISSUE CANNOT BE ENTIRELY RULED OUT AS A CONTRIBUTING FACTOR AS PRECISION TESTING TO ASSESS INSTRUMENT PERFORMANCE WAS NOT PERFORMED BY THE CUSTOMER AT THE TIME OF WRITING THIS REPORT. IN ADDITION, PRE-ANALYTICAL SAMPLE PROCESSING COULD NOT BE RULED OUT AS A CONTRIBUTING FACTOR, AS IT WAS UNKNOWN IF THE CUSTOMER WAS FOLLOWING THE SAMPLE COLLECTION DEVICE MANUFACTURE'S RECOMMENDATION FOR SAMPLE CENTRIFUGATION. GIVEN THAT MULTIPLE ANALYTE RESULTS FROM THE INITIAL TEST EVENT WERE DISCORDANT WHEN COMPARED TO ANY OF THE FIVE REPEAT TEST EVENTS OF THE SAME PATIENT SAMPLE, THE ORTHO TECHNICAL SOLUTIONS CENTER (TSC) INVESTIGATED WHETHER OR NOT A PRE-ANALYTICAL PATIENT SAMPLE MIX UP CONTRIBUTED TO THIS EVENT. HOWEVER, FINDINGS FROM THIS INVESTIGATION USING ECONNECTIVITY WERE INCONCLUSIVE. THEREFORE, IT IS POSSIBLE THAT A PRE-ANALYTICAL PATIENT SAMPLE MIX UP HAD CONTRIBUTED TO THIS EVENT, BUT THIS COULD NOT BE CONFIRMED. CONTINUAL TRACKING AND TRENDING OF COMPLAINTS HAS NOT IDENTIFIED ANY SIGNALS THAT WOULD INDICATE A POTENTIAL SYSTEMATIC ISSUE WITH VITROS PCT REAGENT LOT 0440.

Description of Event or Problem

A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT A NON-REPRODUCIBLE, HIGHER THAN EXPECTED VITROS PROCALCITONIN (PCT) RESULT WAS OBTAINED FROM A SAMPLE FROM A SINGLE PATIENT WHEN TESTED ON A VITROS XT 7600 INTEGRATED SYSTEM USING REAGENT LOT 0440 WHEN COMPARED TO A REPEAT PCT TEST RESULT OBTAINED FROM A REPEAT TEST OF THE SAME SAMPLE ON THE SAME VITROS XT 7600 INTEGRATED SYSTEM. PATIENT SAMPLE RESULT OF 1.916 NG/ML VERSUS THE EXPECTED (REPORTED) REPEAT RESULT OF 0.059 NG/ML BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED, THE NON-REPRODUCIBLE, HIGHER THAN EXPECTED VITROS PCT RESULT OF 1.916 NG/ML WAS QUESTIONED BY THE LABORATORY STAFF AND REPEAT TEST EVENTS WERE SUBSEQUENTLY PROCESSED. THE QUESTIONED RESULT WAS NOT REPORTED FROM THE LABORATORY AND THERE HAS BEEN NO ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBER (B)(4) AND REPORTABILITY ASSESSMENT (B)(4).