MalfunctionQOF

COBAS SARS-COV-2 & INFLUENZA A/B FOR USE ON THE COBAS LIAT SYSTEM

Received Mar 7, 2025 · Event occurred Feb 5, 2025

Report 2243471-2025-00656 · MDR key 21546148

Device

Generic name

Multi-target Resp Specimen Nat Incl. Sars-cov-2 & Other Microbial Agents

Catalog number

09211101190

Lot number

41014L

Product problems

  • Non Reproducible Results
  • Non Reproducible Results

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

PRODUCT ISSUE FOUND/NO SYSTEMIC IMPACT. THE DISCREPANCY OBSERVED IS MOST LIKELY DUE TO INTERFERENCE FROM A HIGH VIRAL LOAD OF FLU A AND A LOW VIRAL LOAD OF SARS. ACCORDING TO THE METHOD SHEET UNDER PROCEDURAL LIMITATIONS, FALSE NEGATIVE RESULTS MAY OCCUR IF A SPECIMEN IS IMPROPERLY COLLECTED, TRANSPORTED, OR HANDLED; IF THERE IS INSUFFICIENT RNA TO BE DETECTED; OR IF ONE OR MORE TARGET VIRUSES INHIBIT THE AMPLIFICATION OF OTHER TARGETS. IT IS POSSIBLE THAT THE VIRAL LOAD OF THE SARS INFECTION WAS LOW AT THE TIME OF COLLECTION.

Additional Manufacturer Narrative

THE SERIAL NUMBER OF THE COBAS LIAT ANALYZER WAS (B)(6). INVESTIGATION IS ONGOING.

Description of Event or Problem

THERE WAS AN ALLEGATION OF A QUESTIONABLE RESULT FOR A SINGLE PATIENT SAMPLE TESTED WITH THE COBAS SARS-COV-2 & INFLUENZA A/B NUCLEIC ACID TEST FOR USE ON THE COBAS LIAT SYSTEM. IT WAS REPORTED THAT THE SAMPLES INITIALLY GENERATED POSITIVE RESULTS FOR INFLUENZA A AND SARS-COV-2 AND NEGATIVE RESULTS FOR INFLUENZA B. THE SAME SAMPLE WAS RETESTED ON A DIFFERENT LIAT ANALYZER AND GENERATED A POSITIVE RESULT FOR INFLUENZA A AND NEGATIVE RESULTS FOR INFLUENZA B AND SARS-COV-2.