MalfunctionNIG

ATELLICA IM CA 19-9

Received Mar 6, 2025 · Event occurred Feb 4, 2025

Report 1219913-2025-00048 · MDR key 21534110

Device

Generic name

Atellica Im Ca 19-9

Model number

N/A

Catalog number

10995490

Lot number

10595541

Product problems

  • Low Test Results
  • Low Test Results

Patient

70 YR · Male

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

A CUSTOMER FROM OUTSIDE THE UNITED STATES REPORTED AN OBSERVATION OF A DEPRESSED ATELLICA IM CA 19 9 (CA 19 9) RESULT WHICH WAS DISCORDANT RELATIVE TO DILUTED RESULT. THE CALIBRATION WAS VALID AND QUALITY CONTROL (QC) RECOVERED WITHIN THE LABORATORY ESTABLISHED RANGES. THE ASSAY'S INSTRUCTIONS FOR USE (IFU) STATES THE FOLLOWING, UNDER INTERPRETATION OF RESULTS: "RESULTS OF THIS ASSAY SHOULD ALWAYS BE INTERPRETED IN CONJUNCTION WITH THE PATIENT'S MEDICAL HISTORY, CLINICAL PRESENTATION, AND OTHER FINDINGS." SIEMENS IS EVALUATING THE EVENT.

Description of Event or Problem

THE CUSTOMER REPORTS OBSERVATION OF A DEPRESSED ATELLICA IM CA 19 9 (CA 19 9) RESULT WHICH WAS DISCORDANT RELATIVE TO THE DILUTED RESULT WHEN USING LOT 541. AN INITIAL DEPRESSED RESULT WAS OBTAINED FOR THE PATIENT SAMPLE IN QUESTION. THE INITIAL RESULT WAS REPORTED TO THE PHYSICIAN WHO QUESTIONED THE RESULT. THE SAMPLE WAS RETESTED ON THE ORIGINAL ANALYZER WHICH PRODUCED A SIMILAR DEPRESSED RESULT. LATER, THE SAME SAMPLE WAS DILUTED (1:10) AND RETESTED ON THE ORIGINAL INSTRUMENT WHICH OBTAINED AN ELEVATED RESULT. THE ELEVATED RESULT WAS CONSIDERED CORRECT AND ISSUED ON THE CORRECTED REPORT TO THE PHYSICIAN. FOR TROUBLESHOOTING PURPOSES, THE SAME SAMPLE WAS TESTED USING SERIAL DILUTION ON THE ORIGINAL ANALYZER AND ELEVATED RESULTS WERE OBTAINED COMPARED TO THE NEAT RESULT. THE SAMPLE WAS ALSO TESTED BOTH NEAT AND DILUTED ON AN ALTERNATE METHOD TESTING AND OBTAINED ELEVATED RESULTS. THERE ARE NO ALLEGATIONS OF ANY PATIENT HARM RESULTING FROM THE OBSERVED RESULT DISCORDANCE.

Additional Manufacturer Narrative

MDR 1219913-2025-00048 WAS INITIALLY FILED ON 06-MAR-2025. ADDITIONAL INFORMATION (07-MAR-2025): SIEMENS CONCLUDED INVESTIGATION FOR A CUSTOMER COMPLAINT OF OBSERVATION OF A DEPRESSED ATELLICA IM CA 19 9 (CA 19 9) RESULT WHICH WAS DISCORDANT RELATIVE TO DILUTED RESULT. SIEMENS INVESTIGATION CONFIRMED THAT AIM CA 19-9 DILUTION RECOVERY WAS CONSISTENT WITH DESIGN VERIFICATION PERFORMANCE. FURTHER REVIEW INDICATED THAT SOME PATIENT SAMPLES EXHIBITED DILUTION OVER-RECOVERY WITH MUCIN/MUCIN-LIKE ASSAYS SUCH AS CA 19-9. CUSTOMERS WERE ADVISED TO REFER TO THE ATELLICA IM IFU NOTE ON 'SAMPLE-DEPENDENT NONLINEAR DILUTIONS,' SUPPORTED BY RELEVANT LITERATURE. THE ASSAY WAS INTENDED FOR MONITORING, NOT SCREENING, WITH RESULTS RANGING FROM 1.20 ¿ 700 U/ML. ONLY SAMPLES EXCEEDING 700 U/ML REQUIRED DILUTION BEFORE RETESTING. A REVIEW OF ASSAY CONTROL SYSTEM MATERIALS CONFIRMED THAT PERFORMANCE REMAINED WITHIN SPECIFICATION OVER THE PAST TWO YEARS. SIEMENS INVESTIGATED CA 19-9 DILUTION-RECOVERY PERFORMANCE ACROSS MULTIPLE REAGENT LOTS, WHICH INCLUDED BOTH MANUAL AND AUTO 1:10 DILUTIONS WITH MEDICAL DECISION POOL (MDP) LEVEL 5, WHICH IS COMMUTABLE TO PATIENT SAMPLES, HIGH CALIBRATOR LOTS C941H, C943H AND C945H AND HIGH DOSE PATIENT SAMPLES. THE OBSERVED PERFORMANCE MET THE ATELLICA IM DESIGN VERIFICATION REQUIREMENTS, WHICH ALLOW A +/- 30% TOLERANCE FOR AUTO-DILUTION RECOVERY. FURTHER INVESTIGATION DEMONSTRATED THAT THE CA 19-9 ASSAY MET ACCURACY SPECIFICATIONS FOR ALL INTERNAL CONTROL MATERIALS TESTED DURING LOT-QUALIFICATION OF LOTS RELEASED IN THE PAST TWO YEARS. IT IS NOTED THAT DILUTION-RECOVERY PERFORMANCE IS NOT CONSISTENT ACROSS ALL SAMPLES, AND IT IS POSSIBLE THAT SOME PATIENT SAMPLES MAY EXHIBIT DILUTION OVER-RECOVERY WITH MUCIN/MUCIN-LIKE ASSAYS SUCH AS CA 19-9. THE PRODUCT¿S INSTRUCTIONS FOR USE (IFU) STATES ¿SAMPLE DEPENDENT NONLINEAR DILUTIONS CAN BE OBSERVED.¿ [REFERENCE: WILD D, ED. THE IMMUNOASSAY HANDBOOK. 4TH ED. OXFORD, UK: ELSEVIER SCIENCE; 2013:842]. ADDITIONALLY, IT IS IMPORTANT TO NOTE THAT THE ATELLICA IM CA19-9 IFU STATES THAT THE ASSAY IS USEFUL AS AN AID IN MONITORING THE STATUS OF PATIENTS HAVING CONFIRMED PANCREATIC CANCER WHO HAVE LEVELS OF SERUM AT SOME POINT EXCEEDING THE MEDIAN CONCENTRATION. THE ASSAY IS NOT INTENDED FOR SCREENING PURPOSES AND SHOULD ALWAYS BE USED IN CONJUNCTION WITH ALL OTHER CLINICAL AND LABORATORY DATA. THE ASSAY PROVIDES RESULTS FROM 1.20 ¿ 700 U/ML. ONLY SAMPLES WITH LEVELS >700 U/ML SHOULD BE DILUTED BEFORE RE-TESTING. BASED ON THE INVESTIGATION, A PRODUCT PROBLEM WAS NOT IDENTIFIED AND NO FURTHER INVESTIGATION IS REQUIRED. THE ISSUE WAS RESOLVED BY ROUTINE TROUBLESHOOTING. CUSTOMER IS OPERATIONAL; NO FURTHER ACTIONS ARE REQUIRED. IN SECTION H6 OF THIS REPORT, THE INVESTIGATION FINDING AND CONCLUSION CODES WERE UPDATED BASED ON THE INVESTIGATION RESULTS.