TRYPTICASE SOY BROTH
Received Mar 4, 2025 · Event occurred Feb 10, 2025
Report MW5167167 · MDR key 21513844
Device
Generic name
Culture Media, Non-selective And Non-differential
Manufacturer
Remel, Inc.Lot number
185090
Patient
67 YR · Male
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
THIS AFFECTED MULTIPLE PATIENTS (86) WHO HAD TISSUE SPECIMEN CULTURES. ONE OF THE MICRO COLLEAGUES REPORTED SEEING A TREND WITH GRAM STAIN RESULTS OF STREPLIKE GRAM-POSITIVE COCCI IN CHAINS IN MULTIPLE SPECIMENS. NONE OF THE CULTURES WERE GROWING ANY ORGANISMS. AFTER AN EXTENSIVE INVESTIGATION, IT WAS DETERMINED THAT THE TRYPTICASE SOY BROTH THAT WE WERE USING TO GRIND THE TISSUE IN WAS CONTAMINATED. TEN BOTTLES OF THE BROTH WAS GRAM STAINED AND STREPLIKE GRAM-POSITIVE COCCI WERE FOUND IN THE TSB (TRYPTICASE SOY BROTH) BROTH. REFERENCE REPORTS: MW5167166, MW5167168, MW5167169, MW5167170, MW5167171, MW5167172, MW5167173, MW5167174, MW5167175.