GUNTHER TULIP VENA CAVA FILTER RETRIEVAL SET

Received Feb 26, 2025 · Event occurred Jan 28, 2025

Report 1820334-2025-00190 · MDR key 21471685

Device

Generic name

Mmx Device, Percutaneous Retrieval

Manufacturer

Cook Inc

Model number

G13287

Catalog number

GTRS-200-RB

Lot number

16121923

Product problems

  • Mechanical Problem
  • Material Split, Cut or Torn
  • Mechanical Problem
  • Material Split, Cut or Torn

Patient

82 YR · Female

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

BLANK FIELDS ON THIS FORM INDICATE THE INFORMATION IS UNKNOWN OR UNAVAILABLE. ADDITIONAL/CORRECTED INFORMATION: B5, D10. THIS REPORT INCLUDES INFORMATION KNOWN AT THIS TIME. A FOLLOW UP REPORT WILL BE SUBMITTED SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE. THIS REPORT IS REQUIRED BY THE FDA UNDER 21 CFR PART 803. THIS REPORT IS BASED ON UNCONFIRMED INFORMATION SUBMITTED BY OTHERS. NEITHER THE SUBMISSION OF THIS REPORT NOR ANY STATEMENT MADE IN IT IS INTENDED TO BE AN ADMISSION THAT ANY COOK DEVICE IS DEFECTIVE OR MALFUNCTIONED; THAT A DEATH OR SERIOUS INJURY OCCURRED; OR THAT ANY COOK DEVICE CAUSED OR CONTRIBUTED TO; OR IS LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY IF A MALFUNCTION OCCURRED.

Description of Event or Problem

INFORMATION WAS AVAILABLE BUT INADVERTENTLY OMITTED FROM THE PREVIOUS REPORT. THE DEVICE WAS USED TO RETRIEVE ANOTHER MANUFACTURER¿S INFERIOR VENA CAVA FILTER THAT WAS PLACED (B)(6) 2024. THE DEVICE APPEARED NORMAL DURING PREPARATION. THERE WAS NO DIFFICULTY ADVANCING, REACHING, OR CAPTURING THE FILTER. BOTH THE BLACK AND BLUE SHEATHS REPORTEDLY RIPPED.

Additional Manufacturer Narrative

SUMMARY OF EVENT: AS REPORTED, DURING ROUTINE RETRIEVAL OF ANOTHER MANUFACTURER'S INFERIOR VENA CAVA FILTER (ORIGINALLY PLACED ON (B)(6) 2024) IN INTERVENTIONAL RADIOLOGY, A GUNTHER TULIP VENA CAVA FILTER RETRIEVAL SET¿S SHEATH "TORE" AND WAS POSSIBLY RIPPED BY THE FILTER. THE DEVICE APPEARED NORMAL DURING PREPARATION. THE FILTER HOOK WAS REPORTEDLY CAPTURED AND SNARED UNDER DIRECT FLUOROSCOPIC VISUALIZATION, AND THE SHEATHS WERE ADVANCED OVER THE FILTER. THE USER NOTED THAT THE SHEATH WAS KINKED WITH THE FILTER INSIDE. THE FILTER WAS REPORTEDLY UNABLE TO BE RE-COLLAPSED; THEREFORE, THE SHEATH AND FILTER WERE REMOVED ENDOVASCULARLY FROM THE PATIENT, WITH "A COUPLE" OF THE FILTER LEGS STICKING OUT FROM THE SIDE OF THE SHEATH. THERE WAS NO DIFFICULTY ADVANCING, REACHING, OR CAPTURING THE FILTER. BOTH THE BLACK AND BLUE SHEATHS REPORTEDLY RIPPED. AFTER REMOVAL, THE USER NOTED THAT THE SHEATH WAS RIPPED AND TORN LONGITUDINALLY. PER THE REPORTER, THE FILTER ¿SHREDDED THE SHEATH¿. THE FILTER, WHICH WAS REMOVED IN ITS ENTIRETY, WAS NOT DAMAGED UPON INSPECTION. A COMPLETION VENOGRAM WAS PERFORMED. THE IVC WAS PATENT, WITHOUT EVIDENCE OF INJURY, EXTRAVASATION, OR THROMBUS. USER ERROR WAS NOT REPORTED. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THERE HAS BEEN NO REPORT THAT THE PATIENT REQUIRED ADDITIONAL PROCEDURES OR EXPERIENCED ANY ADVERSE EFFECTS DUE TO THIS EVENT. CORRECTED INFORMATION: H6 (ANNEXES A & G) INVESTIGATION EVALUATION: REVIEWS OF THE COMPLAINT HISTORY, DEVICE HISTORY RECORD (DHR), INSTRUCTIONS FOR USE (IFU), MANUFACTURING INSTRUCTIONS, AND QUALITY CONTROL PROCEDURES WERE CONDUCTED DURING THE INVESTIGATION. THE COMPLAINT DEVICE WAS NOT RETURNED TO COOK FOR INVESTIGATION. A DOCUMENT-BASED INVESTIGATION EVALUATION WAS PERFORMED. A REVIEW OF THE DEVICE HISTORY RECORD FOUND NO NON-CONFORMANCES ON THE LOT. A REVIEW OF COMPLAINT HISTORY FOUND NO ADDITIONAL COMPLAINTS FOR THIS LOT NUMBER. THE PRODUCT IFU STATES THAT THE DEVICE IS INTENDED FOR RETRIEVAL OF COOK GUNTHER TULIP AND CELECT FILTERS. A REVIEW OF THE DEVICE MASTER RECORD (DMR) CONCLUDED THAT SUFFICIENT INSPECTION ACTIVITIES ARE IN PLACE TO IDENTIFY THIS FAILURE MODE PRIOR TO DISTRIBUTION. THE INFORMATION PROVIDED UPON REVIEW OF THE DMR, DHR, AND IFU SUGGESTS THAT THERE IS EVIDENCE THE DEVICE WAS MANUFACTURED TO SPECIFICATION. THERE IS NO EVIDENCE OF NON-CONFORMING DEVICES IN-HOUSE OR IN THE FIELD. BASED ON THE INFORMATION PROVIDED AND THE RESULTS OF THE INVESTIGATION, COOK HAS CONCLUDED THAT OFF-LABEL USE CONTRIBUTED TO THIS EVENT. THE GUNTHER TULIP VENA CAVA FILTER RETRIEVAL SET IS INTENDED FOR RETRIEVAL OF IMPLANTED COOK GUNTHER TULIP AND COOK CELECT VENA CAVA FILTERS IN PATIENTS WHO NO LONGER REQUIRE A FILTER. IN THIS CASE, THE DEVICE WAS USED TO RETRIEVE A NON-COOK FILTER. THE RISK ANALYSIS FOR THIS FAILURE MODE WAS REVIEWED AND NO ADDITIONAL ESCALATION WAS REQUIRED. THE APPROPRIATE PERSONNEL HAVE BEEN NOTIFIED, AND COOK WILL CONTINUE TO MONITOR FOR SIMILAR EVENTS. THIS REPORT IS REQUIRED BY THE FDA UNDER 21 CFR PART 803 AND IS BASED ON UNCONFIRMED INFORMATION SUBMITTED BY OTHERS. NEITHER THE SUBMISSION OF THIS REPORT NOR ANY STATEMENT CONTAINED HEREIN IS INTENDED TO BE AN ADMISSION THAT ANY COOK DEVICE IS DEFECTIVE OR MALFUNCTIONED, THAT A DEATH OR SERIOUS INJURY OCCURRED, NOR THAT ANY COOK DEVICE CAUSED, CONTRIBUTED TO, OR IS LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY IF A MALFUNCTION OCCURRED.

Description of Event or Problem

AS REPORTED, DURING ROUTINE RETRIEVAL OF AN UNKNOWN INFERIOR VENA CAVA (IVC) FILTER IN INTERVENTIONAL RADIOLOGY, A GUNTHER TULIP VENA CAVA FILTER RETRIEVAL SET¿S SHEATH ¿TORE¿ AND WAS POSSIBLY RIPPED BY THE UNKNOWN FILTER. THE FILTER HOOK WAS REPORTEDLY CAPTURED AND SNARED UNDER DIRECT FLUOROSCOPIC VISUALIZATION, AND THE SHEATHS WERE ADVANCED OVER THE FILTER. THE USER NOTED THAT THE SHEATH WAS KINKED WITH THE FILTER INSIDE. THE FILTER WAS REPORTEDLY UNABLE TO BE RE-COLLAPSED; THEREFORE, THE SHEATH AND FILTER WERE REMOVED ENDOVASCULARLY FROM THE PATIENT, WITH ¿A COUPLE¿ OF THE FILTER LEGS STICKING OUT FROM THE SIDE OF THE SHEATH. AFTER REMOVAL, THE USER NOTED THAT THE SHEATH WAS RIPPED AND TORN LONGITUDINALLY. PER THE REPORTER, THE FILTER ¿SHREDDED THE SHEATH¿. THE FILTER, WHICH WAS REMOVED IN ITS ENTIRETY, WAS NOT DAMAGED UPON INSPECTION. A COMPLETION VENOGRAM WAS PERFORMED. THE IVC WAS PATENT, WITHOUT EVIDENCE OF INJURY, EXTRAVASATION, OR THROMBUS. USER ERROR WAS NOT REPORTED. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT¿S BODY. THERE HAS BEEN NO REPORT THAT THE PATIENT REQUIRED ADDITIONAL PROCEDURES OR EXPERIENCED ANY ADVERSE EFFECTS DUE TO THIS EVENT.

Additional Manufacturer Narrative

E3: OCCUPATION = SENIOR PORTFOLIO MANAGER CV/IR. H3: DEVICE EVALUATION ANTICIPATED, BUT NOT YET BEGUN. THIS REPORT INCLUDES INFORMATION KNOWN AT THIS TIME. A FOLLOW-UP REPORT WILL BE SUBMITTED SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE. THIS REPORT IS REQUIRED BY THE FDA UNDER 21 CFR PART 803 AND IS BASED ON UNCONFIRMED INFORMATION SUBMITTED BY OTHERS. NEITHER THE SUBMISSION OF THIS REPORT NOR ANY STATEMENT CONTAINED HEREIN IS INTENDED TO BE AN ADMISSION THAT ANY COOK DEVICE IS DEFECTIVE OR MALFUNCTIONED, THAT A DEATH OR SERIOUS INJURY OCCURRED, NOR THAT ANY COOK DEVICE CAUSED, CONTRIBUTED TO, OR IS LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY IF A MALFUNCTION OCCURRED.