Received Jun 28, 2011 · Event occurred Jun 14, 2011

Report 1823260-2011-03499 · MDR key 2143257

Device

Generic name

Biuret (colorimetric), Total Protein

Manufacturer

Roche Diagnostics

Model number

NA

Catalog number

12000903216

Lot number

63476401

Product problems

  • High Test Results

Patient

Not reported

  • No Known Impact Or Consequence To Patient

Narrative

Description of Event or Problem

THE USER RECEIVED A QUESTIONABLE TOTAL PROTEIN RESULT FOR ONE PATIENT SAMPLE FROM THE ANALYTICAL P MODULE ANALYZER SERIAL NUMBER (B)(4). THE INITIAL RESULT FROM AN ALIQUOT OF THE SAMPLE WAS 11.8 G/DL AND WAS REPORTED OUTSIDE THE LABORATORY. THE ORIGINAL SAMPLE WAS REPEATED AND THE RESULT WAS 6.6 G/DL. THE ALIQUOT WAS THEN REPEATED AND THE RESULT WAS 6.6 G/DL. THE REPORTED RESULT WAS CORRECTED ON (B)(6) 2011. THE USER DID NOT HAVE ACCESS TO INFORMATION CONCERNING IF THE PATIENT WAS ADVERSELY AFFECTED. THE TOTAL PROTEIN R2 REAGENT LOT NUMBER WAS 63554801. UPON EVALUATION OF THE ANALYZER, THE FIELD SERVICE REPRESENTATIVE COULD NOT DETERMINE A CAUSE. HE CHECKED MULTIPLE AREAS OF THE ANALYZER AND ALL WERE ACCEPTABLE. TO VERIFY THE ANALYZER OPERATION, HE RAN PRECISION TESTING ON THE TOTAL PROTEIN ASSAY.