COBAS AMPLICOR CHLAMYDIA TACHOMATIS TEST
Received Jun 28, 2011 · Event occurred Mar 15, 2011
Report 2243471-2011-00069 · MDR key 2142254
Device
Generic name
Dna Probe, Nucleic Acid Amplification, Chlamydia
Manufacturer
Roche Molecular SystemsCatalog number
20757497122
Product problems
- False Negative Result
Patient
Not reported
- No Information
Narrative
Description of Event or Problem
A CUSTOMER SITE IN (B)(6) ALLEGED THAT FALSE (B)(6) RESULTS WERE GENERATED DURING A COMPARISON STUDY WITH THE COBAS AMPLICOR CHLAMYDIA TRACHOMATIS TEST. (B)(4) THEY RAN PREVIOUSLY RUN (B)(6) SAMPLES ON THREE OF THEIR INSTRUMENTS AND ONE OF THE INSTRUMENTS - (B)(4)- GENERATED DISCREPANT RESULTS. THE SAMPLES HAD BEEN STORED AT 2-8 DEGREES FOR ONE DAY THEN WERE FROZEN AT -20 DEGREES.
Additional Manufacturer Narrative
(B)(4). NO FAILURE DETECTED AND PRODUCT WITHIN SPECIFICATION.URINE SAMPLES FROM TWO PATIENTS PREVIOUSLY GENERATED (B)(4) CT RESULTS. UPON RETESTING FOR A COMPARISON STUDY, (B)(4) RESULTS WERE OBTAINED FOR SOME REPLICATES. THE SAMPLES HAD BEEN STORED AT 2-8C FOR ONE DAY AND THEN WERE FROZEN AT -20C. THE CUSTOMER REPORTED THAT THE PATIENTS WHOSE SAMPLES PRODUCED THE DISCREPANT RESULTS WERE NOT ADVERSELY AFFECTED.THE CUSTOMER WAS NOT WILLING TO PROVIDE ANY FURTHER INFORMATION FOR THE INVESTIGATION, INCLUDING KIT LOT NUMBERS OR LABORATORY STATISTICS (B)(4). NO PATIENT SAMPLES WERE PROVIDED FOR INVESTIGATIVE TESTING.BASED ON THE LIMITED AVAILABLE INFORMATION, NO PRODUCT NON-CONFORMANCE WAS IDENTIFIED. (B)(4). IT IS ALSO NOTED THAT THE SENSITIVITY WAS LOWER FOR FROZEN MALE URINE SPECIMENS AND HIGHER FOR FROZEN FEMALE URINE SPECIMENS, AS COMPARED TO SPECIMENS WHICH HAD NOT BEEN FROZEN. IT IS UNKNOWN WHETHER THE TWO SAMPLES WHICH GENERATED DISCREPANT RESULTS WERE MALE OR FEMALE. THE CUSTOMER DID NOT PROVIDE ANY STATISTICS THAT COULD BE USED TO DETERMINE IF THE DISCREPANT RESULTS GENERATED WERE WITHIN PRODUCT LABELING CLAIMS.(B)(4).
Additional Manufacturer Narrative
A DEFINITIVE CONCLUSION CANNOT BE DRAWN AT THIS TIME, AS THE INVESTIGATION INTO THIS ISSUE IS ONGOING. THE OUTCOME OF THE INVESTIGATION WILL BE COMMUNICATED THROUGH A FOLLOW-UP REPORT. (B)(4).