Received Feb 20, 2025 · Event occurred Jan 30, 2025

Report 1824206-2025-00363 · MDR key 21421575

Device

Generic name

Bed, Flotation Therapy, Powered

Model number

P6881

Catalog number

P6881

Product problems

  • Material Split, Cut or Torn
  • Material Split, Cut or Torn

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN STATING THE POWER CORD HAD DAMAGE WITH EXPOSED COPPER WIRES. THE BED WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED.

Additional Manufacturer Narrative

THE BAXTER TECHNICIAN FOUND THE POWER CORD NEEDED TO BE REPLACED. PER THE HILLROM SERVICE MANUAL INSPECT POWER CORD FOR DAMAGE, TWISTING AND REPLACE IF NECESSARY. A SEARCH OF THE BAXTER MAINTENANCE RECORDS DID NOT SHOW BAXTER PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNKNOWN IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR BEDS. THE TECHNICIAN REPLACED THE POWER CORD TO RESOLVE THE REPORTED EVENT. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.