PENTAX
Received Feb 20, 2025 · Event occurred Feb 14, 2025
Report 9610877-2025-51027 · MDR key 21418042
Device
Generic name
Fiberopticintubationscope 1.2c 3.0tp Rop
Manufacturer
Hoya Corporation Pentax Tokyo OfficeModel number
FI-9RBSProduct problems
- Fluid/Blood Leak
- Fluid/Blood Leak
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE (FLUID DAMAGE).
Additional Manufacturer Narrative
G4: THIS DEVICE IS NOT DISTRIBUTED IN US SO THAT 510K# IS BLANK. D4 UDI: "NOT DISTRIBUTED IN USA", BECAUSE PRODUCTS ARE NOT DISTRIBUTED IN USA PRODUCTS, BUT SIMILAR DEVICE PRODUCT ARE LISTED IN USA. THE PRODUCT WAS RETURNED TO PENTAX MEDICAL FOR REPAIR. OUR TECHNICIAN CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE CFB (COHERENT FIBER BUNDLE) FLUID DAMAGE. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE FLUID DAMAGE FROM THE CFB (COHERENT FIBER BUNDLE). IN ADDITION, OUR TECHNICIAN CONFIRMED THAT THE CONTROL BODY FLUID DAMAGE, THE FORWARD BODY FRAME FLUID DAMAGE, THE FORWARD BODY COVER FLUID DAMAGE, THE BIOPSY INLET T-PIECE FLUID DAMAGE, THE LIGHT SOURCE CONNECTOR FLUID DAMAGE, THE INSERTION FLEXIBLE TUBE PERFORATED, AND THE BRAND PLATE WORN OUT; HOWEVER, THESE DEFECTS ARE NOT THE MAIN CAUSE, AND/OR IRRELEVANT TO THE ALLEGED COMPLAINT. BASED ON THE TECHNICAL REPORT ""HR-RPT-0586(IMAGE FAILURE)"" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.