OR1 NEO IP, ENH2
Received Feb 19, 2025 · Event occurred Jan 29, 2025
Report 9610617-2025-00234 · MDR key 21411340
Device
Generic name
Or1 Neo Ip, Enh2
Manufacturer
Karl Storz Se & Co. KgModel number
WUIS2548-01Catalog number
WUIS2548-01
Product problems
- Output Problem
- Output Problem
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
IT HAS BEEN REPORTED THAT THE INSTALL ON THIS ROOM LAST WEEK AND ALL THREE VIDEO PLATES ON THE MONITOR ARM ARE NOT PASSING A SIGNAL TO THE NEW IP. IT OCCURED DURING A PROCEDURE WITH NO PATIENT IMPACT. A SECONDARY NON-NETWORKS VIDEO DEVICES HAD TO BE MOVED INTO THE OR TO COMPLETE THE PROCEDURE. FORTUNATELY, IT WAS IN AN INTERVENTIONAL SUITE WHERE THE INTERVENTIONAL IMAGE WAS NOT INTERRUPTED INDEPENDENT FROM OUR SYSTEM. THE NAVIGATION MONITOR HAD TO BE MOVED OVER AS THE CATHETER WAS IN THE PATIENT DURING THE PROCEDURE IN ORDER FOR THE PHYSICIAN TO SEE AT LOCATION WHERE HE WAS RELYING ON OUR MONITOR NARRATION TO NAVIGATE.
Additional Manufacturer Narrative
THE DEVICE WILL BE FORWARDED TO THE MANUFACTURING SITE FOR INVESTIGATION. SHOULD RELEVANT ADDITIONAL INFORMATION / INVESTIGATION RESULTS BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED UNSOLICITED. THIS EVENT IS FILED UNDER INTERNAL COMPLAINT ID (B)(4).
Additional Manufacturer Narrative
CORRECTIONS MADE IN SECTIONS: D2B, D4, E1, G4, AND H11. THE DEVICE WILL NOT BE FORWARDED TO THE MANUFACTURING SITE FOR INVESTIGATION. RATHER, THEY WILL SEND A SERVICE TECH TO EVALUATE THE DEVICE IN THE FIELD. SHOULD RELEVANT ADDITIONAL INFORMATION / INVESTIGATION RESULTS BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED UNSOLICITED. THIS EVENT IS FILED UNDER INTERNAL COMPLAINT ID (B)(4).
Additional Manufacturer Narrative
THIS PRODUCT IS MANUFACTURED AT KARL STORZ ENDOSCOPY AMERICA, 13803 N PROMENADE BLVD, STAFFORD, TX 77477; HOWEVER, IT IS DESIGNED IN GERMANY. DEVICE EVALUATION: DEFECTIVE FIBER TERMINATION AT NETWORK SWITCH AND POTENTIAL POWER INSTABILITY FROM FAULTY UPS UNIT (POWERVAR THIRD-PARTY DEVICE). BOTH ISSUES CONTRIBUTED INTERMITTENT SIGNAL LOSS ON VIDEO PLATES. THE ISSUE WAS RESOLVED BY FIXING THE FIBER TERMINATION, REPLACING THE UPS, AND BOTH VIDEO PLATES. NO FURTHER ISSUES NOTED. THE EVENT IS FILED UNDER INTERNAL KARL STORZ COMPLAINT ID: (B)(4).