PROACCESS(TM) VERTICAL MAST
Received Feb 18, 2025 · Event occurred Jan 29, 2025
Report 8030965-2025-01672 · MDR key 21401102
Device
Product problems
- Mechanical Problem
- Mechanical Problem
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. H3, H4, H6 THE PRODUCT WAS RETURNED TO J&J MEDTECH ORTHOPAEDICS FOR EVALUATION. VISUAL INSPECTION OF THE RETURNED DEVICE FOUND NOTHING INDICATIVE OF A DEVICE NONCONFORMANCE. A FUNCTIONAL TEST WAS PERFORMED AND THE ATTACHING MECHANISM OF THE COUPLER IS RETAINED SECURELY AT THE TUBE. A DIMENSIONAL INSPECTION WAS NOT PERFORMED SINCE IT WAS NOT APPLICABLE TO THE COMPLAINT CONDITION. THE OVERALL COMPLAINT WAS NOT CONFIRMED AS THE OBSERVED CONDITION OF THE PROACCESS(TM) VERTICAL MAST WOULD HAVE NOT CONTRIBUTED TO THE COMPLAINED DEVICE ISSUE. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE REPORTED COMPLAINT CONDITION. AS PART OF J&J MEDTECH ORTHOPAEDICS QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE HISTORY REVIEW: PART: 03.612.001. LOT: 1461752. MANUFACTURING SITE: WERK HAGENDORF. RELEASE TO WAREHOUSE DATE: 27 JUL 2006. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED ARTICLE LOT AND NO NON-CONFORMANCES WERE IDENTIFIED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
Description of Event or Problem
IT WAS REPORTED THAT THE SURGEON FOUND THAT THE POST WAS NOT SECURELY HOLDING THE POST, IT WAS 'WOBBLY'. THERE WAS NO SURGICAL DELAY AS OTHER SET WAS USED. PATIENT WAS NOT INVOLVED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED.
Additional Manufacturer Narrative
PRODUCT COMPLAINT#: (B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803 (AND/OR PART 4, AS APPLICABLE). THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. H3, H6: PRODUCT INVESTIGATION: THE PRODUCT WAS NOT RETURNED TO MEDTECH ORTHOPEDICS; HOWEVER, PHOTOS WERE PROVIDED FOR REVIEW. THE DEVICE ASSOCIATED WITH THIS REPORT WAS NOT RETURNED TO MEDTECH ORTHOPEDICS FOR EVALUATION. THE PHOTOGRAPHS PROVIDED WERE REVIEWED, HOWEVER IN THE IMAGES PROVIDED NO OBSERVATIONS PERTAINING TO THE NATURE OF THE REPORTED EVENT OF LOOSE COULD BE IDENTIFIED. FUNCTIONALITY ISSUES CANNOT BE EVALUATED THROUGH A PHOTO INVESTIGATION. SINCE THE DEVICE WAS NOT RETURNED, A DIMENSIONAL INSPECTION CANNOT BE PERFORMED. THE OVERALL COMPLAINT WAS UNCONFIRMED AS THE PHOTOGRAPHS ATTACHED DON'T HAVE ANY EVIDENCE TO DRAW A CONCLUSION ABOUT THE REPORTED EVENT. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE COMPLAINT CONDITION. AS PART OF MEDTECH ORTHOPEDICS QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. H3, H4, H6: DEVICE HISTORY RECORD (DHR) REVIEW CONDUCTED: PART#: 03.612.001, SYNTHES LOT#: 1461752, SUPPLIER LOT#: 1461752, RELEASE TO WAREHOUSE DATE: 03.AUG.2006. SUPPLIER: (B)(4). NO NCRS WERE GENERATED DURING PRODUCTION. REVIEW OF THE DEVICE HISTORY RECORD(S) SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THIS PRODUCT, AND ANY SUB-COMPONENTS, WHICH WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
Additional Manufacturer Narrative
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. H11 ADDITIONAL NARRATIVE: H3, H6: THE PRODUCT WAS NOT RETURNED TO MEDTECH ORTHOPEDICS ; HOWEVER, PHOTOS WERE PROVIDED FOR REVIEW. ¿ THE DEVICE ASSOCIATED WITH THIS REPORT WAS NOT RETURNED TO MEDTECH ORTHOPEDICS FOR EVALUATION. THE PHOTOGRAPHS ATTACHED WERE REVIEWED, HOWEVER IN THE IMAGES PROVIDED NO OBSERVATIONS PERTAINING TO THE NATURE OF THE REPORTED EVENT OF LOOSE COULD BE IDENTIFIED. FUNCTIONALITY ISSUES CANNOT BE EVALUATED THROUGH A PHOTO INVESTIGATION. SINCE THE DEVICE WAS NOT RETURNED, A DIMENSIONAL INSPECTION CANNOT BE PERFORMED. THE OVERALL COMPLAINT WAS UNCONFIRMED AS THE PHOTOGRAPHS ATTACHED DON¿T HAVE ANY EVIDENCE TO DRAW A CONCLUSION ABOUT THE REPORTED EVENT. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE COMPLAINT CONDITION. AS PART OF MEDTECH ORTHOPEDICS QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE HISTORY REVIEW PART# 03.612.001 SYNTHES LOT# 1461752 SUPPLIER LOT# 1461752 RELEASE TO WAREHOUSE DATE: 03.AUG.2006. SUPPLIER: (B)(4) NO NCRS WERE GENERATED DURING PRODUCTION. REVIEW OF THE DEVICE HISTORY RECORD(S) SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THIS PRODUCT, AND ANY SUB-COMPONENTS, WHICH WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.