MalfunctionQJZ

XLUNG KIT 230

Received Feb 18, 2025 · Event occurred Sep 5, 2024

Report 3012172416-2025-00020 · MDR key 21398297

Device

Generic name

Extracorporeal System For Long-term Respiratory / Cardiopulmonary Failure

Manufacturer

Xenios Ag

Catalog number

32000014

Lot number

8016330

Product problems

  • Incorrect, Inadequate or Imprecise Result or Readings
  • Output Problem
  • Incorrect, Inadequate or Imprecise Result or Readings
  • Output Problem

Patient

NA · Male

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

PLANT INVESTIGATION: THE DESCRIBED SITUATION IS ADEQUATELY ADDRESSED IN IFU AND/OR ON THE LABEL. QUALITY EVENTS MENTIONED IN THE RELEASE CERTIFICATE DO NOT REFER TO THE FAILURE PATTERN AT HAND. NON-CONFORMITY WAS NOT OBSERVED DURING THE MANUFACTURING PROCESS. THE COMPLAINT SAMPLE WAS RECEIVED ON 4/OCT/2024 FOR PRODUCT INVESTIGATION. VISUAL INSPECTION OF THE PUMP HEAD REVEALED SCRATCH MARKS AT THE BASE OF THE INNER HOUSING. DURING FUNCTIONAL TESTING, THE ROTOR OF THE PUMP HEAD WAS TILTED TO ONE SIDE AND THE BASE OF THE ROTOR TOUCHED THE INNER HOUSING. THE ROTOR WAS ROTATING, HOWEVER, THE SOUND OF THE ROTOR SCRAPING AGAINST THE HOUSING AND A SCRAPING NOISE WAS AUDIBLE. THE PUMP HEAD WAS OPENED. SCRATCH MARKS ON THE BASE OF THE INNER HOUSING. IT WAS OBSERVED THAT THE PIN / CALOTTE WAS DAMAGED. THE BALL BEARING OF THE ROTOR WAS NOT DAMAGED; HOWEVER, AN EXCESS OF ADHESIVE WAS OBSERVED AROUND THE CERAMIC BALL AT THE TIP OF THE ROTOR. THIS COULD HAVE LED TO AN UNEVEN RUN OF THE ROTOR AND CAUSED DAMAGE TO THE PIN /CALOTTE DURING OPERATION. IT IS LIKELY THAT AN ERROR OCCURRED DURING GLUING OF THE CERAMIC BALL INTO THE ROTOR AND TOO MUCH GLUE ACCUMULATED AROUND THE CERAMIC BALL AT THE TIP OF THE ROTOR. THIS COULD HAVE CAUSED AN UNEVEN RUN AND DAMAGE TO THE PIN. ANOTHER CAUSE MAY BE AN INITIAL DEFECT IN THE HOUSING (PIN). HOWEVER, THIS CAN NO LONGER BE DETERMINED DUE TO THE DAMAGE.

Description of Event or Problem

A USER FACILITY REPORTED THAT AN ABNORMAL SOUND WAS NOTED AT THE PUMP HEAD OF THE XLUNG KIT 230 DURING EXTRACORPOREAL MEMBRANE OXYGENATION (ECMO) SUPPORT. THE PROVIDED VIDEO EVIDENCE SHOWED ACTIVE ECMO SUPPORT AND DEMONSTRATED A GRINDING SOUND WITH NO OBVIOUS VISIBLE DEFECT TO THE PUMP HEAD OR PUMP DRIVE. UPON FOLLOW UP, IT WAS REPORTED THE NOISE WAS COMING FROM THE PUMP HEAD APPROXIMATELY 4 DAYS AFTER THE START OF SUPPORT. IT WAS CONFIRMED THE PUMP HEAD WAS PROPERLY SEATED IN THE PUMP DRIVE PER INSTRUCTION. THERE WAS NO NOISE DURING THE PRIMING PHASE. IT WAS NOTED THAT THE ROTOR OF THE HORIZONTALLY PLACED PUMP HEAD COULD NOT REMAIN STATIONARY. ECMO SUPPORT WAS CONTINUED WITH A DIFFERENT KIT. THE PATIENT HAD A BLOOD LOSS OF APPROXIMATELY 50 ML AS A RESULT. THE PATIENT DID NOT EXPERIENCE A SERIOUS INJURY OR REQUIRE MEDICAL INTERVENTION AS A RESULT OF THIS EVENT. THE SAMPLE PUMP HEAD WAS RETURNED FOR PRODUCT INVESTIGATION.