inforMED
MalfunctionFSO

OPMI LUMERA 700

Received Feb 14, 2025 · Event occurred Jan 15, 2025

Report 9615010-2025-00002 · MDR key 21387786

Device

Generic name

Opmi Lumera 700

Product problems

  • Unstable
  • Detachment of Device or Device Component
  • Unstable
  • Detachment of Device or Device Component

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THAT DURING A SURGICAL CASE, A MECHANISM DETACHED DUE TO A LOOSE SCREW, CAUSING THE OPMI HEAD OF A LUMERA 700 TO HANG LOOSELY AT A 90 DEGREE ANGLE, ALMOST HITTING THE PATIENT IN THE FACE. THE USER MOVED THE PATIENT TO ANOTHER OPERATING ROOM AND COMPLETED THE CASE. THERE WAS NO INJURY OR NEGATIVE IMPACT TO A PATIENT, USER OR A THIRD PARTY.