PKG, 5MM PEEK MONOPOLAR HANDLE
Received Feb 12, 2025 · Event occurred Jan 24, 2025
Report 0002936485-2025-00109 · MDR key 21369405
Device
Generic name
Endoscopic Grasping/cutting Instrument, Non-powered
Manufacturer
Stryker Endoscopy-san JoseCatalog number
0250080617
Lot number
2441192
Product problems
- Arcing of Electrodes
- Electrical Overstress
- Arcing of Electrodes
- Electrical Overstress
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THIS COMPLAINT INVESTIGATION WAS CLOSED BASED ON THE DEVICE NOT RECEIVED, THEREFORE THE REPORTED FAILURE MODE WAS NOT CONFIRMED. IN THE EVENT THAT THE DEVICE IS RECEIVED, THE COMPLAINT WILL BE REOPENED AND THE INVESTIGATION WILL BE UPDATED WITH NEW RESULTS. ALLEGED FAILURE: ARCING PROBABLE ROOT CAUSE: POOR AUTOCLAVE RELIABILITY CONTACT FORCES USE ERROR INSULATION DAMAGE OR INADEQUATE INSULATION FROM MANUFACTURING/ SERVICING THE REPORTED FAILURE MODE WILL BE MONITORED FOR FUTURE REOCCURRENCE.
Description of Event or Problem
IT WAS REPORTED THAT THE DEVICE WAS ARCING.
Additional Manufacturer Narrative
ADDITIONAL INFORMATION WILL BE PROVIDED ONCE THE INVESTIGATION HAS BEEN COMPLETED. THE DEVICE MANUFACTURER DATE IS NOT KNOWN AT THIS TIME. HOWEVER, SHOULD IT BECOME AVAILABLE IT WILL BE PROVIDED IN FUTURE REPORTS.
Description of Event or Problem
IT WAS REPORTED THAT THE DEVICE WAS ARCING.